Step 1. Capturing reference measurements.
Connect the
SC
pigtail end connector to the
PROLITE-105
output
connector.
Connect the
FC
end of the pigtail to the optical input of the
HD
RANGER
+
(see figure below).
Turn on the
PROLITE-105
and the
HD
RANGER
+
.
In the
HD
RANGER
+
, from the
SPECTRUM
or
MEASUREMENT
screen, press
key to access the
Tools
menu.
Select the
FIBRE OPTIC TEST
option and press the
joystick
.
In the
PROLITE-105
, press once the
SEQ
key to select the
SIMULTANEOUS
mode. This mode simultaneously sends three
wavelengths signals.
In the
HD
RANGER
+
, press the
key
Set Reference
. All current
values are captured, which will be used as reference values.
Now, user can proceed to
Stage 2
in order to measure the attenuation
at each user's access point.