![Primes BeamMonitor BM+ 100S Original Instructions Manual Download Page 33](http://html1.mh-extra.com/html/primes/beammonitor-bm-100s/beammonitor-bm-100s_original-instructions-manual_1601729033.webp)
BeamMonitor BM+
35,0(6
33
Revision 02 EN - 02/2022
Option
Explanation
z-Increment in mm
This option automatically defines the position of the next measurement
on the z-axis (see chapter chapter 9.4.6 „Manual caustic measure“ on
page 39).
Since the BeamMonitor BM+ does not have a z-axis, the laser or the
device must be moved according to the entered value.
X
Enter the z increment in mm.
Manual z-position in mm
This option defines the position of the next measurement on the z-axis
(see chapter 9.4.6 „Manual caustic measure“ on page 39).
Since the BeamMonitor BM+ does not have a z-axis, the laser or the
device must be moved according to the entered value.
X
Enter the manual z-position in mm.
Power P in W *
To calculate the power density, the laser power used must be entered.
Otherwise, the measured amplitudes are given directly in counts.
X
Enter the laser power used during measurement.
autom. Measurement window
If this option is enabled, then the measurement window size will be set
automatically.
X
Set the check mark to enable the option.
Window size in mm
If the
autom. Measurement window
option is disabled, the size of the
measurement window can be set manually. Use one of the following
options:
X
Enter the length and width in the corresponding fields.
X
Position the mouse pointer anywhere within the measurement
range and drag while holding down the left mouse button. Keep the
mouse button pressed until the measurement window meets your
requirements.
Position in mm
Use one of the following options to adjust the position of the measure-
ment window:
X
Enter the x-position / y-position in the corresponding fields.
X
Position the mouse pointer within the measurement window. Then
drag while holding down the right mouse button.
autom. Gain
If this option is enabled, the gain is set automatically.
X
Set the check mark to enable the option.
Gain in dB
If the
autom. Gain
option is disabled, the gain can be set manually.
The option can be used to control the sensitivity of the detector.
Use one of the following options:
X
Enter a value in the input field.
X
Use the slider below the input field.
Time series
A time series consists of several single plane measurements with the
same settings.
Number of measurements *
X
Enter the desired number of single plane measurements of the time
series.
Interval between measurements in s *
X
Enter the pause between the single plane measurements of the time
series. This is the time gap between the end of one measurement
and the start of the next.
Find beam
This option enables an automatic beam search with an automated mea-
suring window size and measurement. The laser beam is automatically
searched for in the entire measuring range. Measurement window size
and gain are set automatically.
The measured plane is then displayed in the graphic display.
X
Click on the button to start the beam search.
Note that the determined measurement data will not be saved in the
project tree of the
Projects
tab.
Tab. 9.1:
Options in the
Device control > Settings
tab of the
Single plane
measuring mode