BeamMonitor BM+
35,0(6
50
Revision 02 EN - 02/2022
Option
Explanation
Line width in mm
Use one of the following options to adjust the line width (length) and
position:
X
Enter the width (length) of the line.
X
Position the mouse pointer at any point within the measurement
area and drag while holding down the left mouse button. Keep the
mouse button pressed until the length of the measuring line meets
your requirements.
Gain in dB
The option can be used to control the sensitivity of the detector.
Use one of the following options:
X
Enter a value in the input field.
X
Use the slider below the input field.
Measuring duration in s
X
Enter the duration of the measurement in s.
# Pixel in x
The number of pixels determines the resolution of the measurement.
Use one of the following options:
X
Enter a value in the input field.
X
Use the slider below the input field.
Parameter settings
All settings in the
Device control
menu can be individually saved for
each device. The saving location is the local installation of the LDS.
These and other options for saving / loading configurations are de-
scribed in chapter 9.3.5 „Saving options“ on page 30.
Save current parameters
X
Click this button to save all current settings of the connected de-
vice.
Load last parameters
X
Click this button to load the last saved device configuration.
Find beam
This option enables an automatic beam search with an automated mea-
suring window size and measurement. The laser beam is automatically
searched for in the entire measuring range. Measurement window size
and gain are set automatically.
The measured plane is then displayed in the graphic display.
X
Click on the button to start the beam search.
Note that the determined measurement data will not be saved in the
project tree of the
Projects
tab.
Power P in W *
To calculate the power density, the laser power used must be entered.
Otherwise, the measured amplitudes are given directly in counts.
X
Enter the laser power used during measurement.
Tab. 9.5:
Options in the
Device control > Settings
tab of the
Linescan