2/2
B 302
15 pin Interface
GB 04543 - Edition 02 - March 12
Analog Output
2 adjustable analog outputs. For each, user can set its allocation
and, if necessary, the start decade (scale start). There is also
1 fixed output (19-ground) for the exponent.
After modification, we advise to save all inputs/outputs set.
Possible values for each output:
Value
Function
Mantissa
1/10 V (*)
Exponent
0/10 V – Low decade to set (*)
Logarithmic
0/10 V - Low decade to set (*)
Inlet pressure
If a pressure measurement gauge is installed in the
customer installation (at the customer’s charge) (*)
2,5 V/8,5 V (10
-3
to 10
+3
mbar)
He compound
0/10 V (compound exponent, mantissa)
(*) Graphs and formulas in appendix: