GB 04527 - Edition 02 - June 12
1/9
C 405
Configuration Menu
Purpose
Setting of the leak detector use
parameters:
hour / date / unit / language,
function keys (
C 200
),
applicative windows,
screeen parameters,
password management,
user levels management.
The update of these parameters is automatically requested during
the first leak detector starting up by the operator: the operator can
modify them at any time afterward.
The set points set are not automatically converted in the new unit in
case of unit change: the operator must to update them.
The hour is not automatically updated at the passage of the summer
time to the winter time and conversely: the operator must to update
them
Hour / Date /
Unit / Language
The function keys allow to activate/stop
a function or to set set points from the
different applicative windows.
Function keys
3 configurations are predefined.
The operator can also define its own function keys configuration.
adixen Vacuum Products - ASM 380 Operating instructions