3/16
A 400
Accessories
GB 03329 - Edition 06 - February 12
Manual isolation valve. The
isolation valve is mandatory for
values greater than 8 · 10
-6
mbar l/s
Temperature coefficient 3 % per degree Celsius for any value
< 8
·
10
-6
mbar l/s (permeation type), 0.2 % per °C for greater
value (capillary type).
Loss 2 %per year for any value < 8
·
10
-6
mbar l/s.
When a specific value is requested, due to the technology used,
The manufacturer can only guarantee an effective nominal value
within ± 20 % of the requested nominal value.
Ex: requested value = 1.8
·
10
-6
mbar l/s
- guaranted delivered value between 1.44 et 2.16
·
10
-6
mbar l/s
- code in DN25 with a valve
FV4611
, and indicates
"requested value = 1.8
·
10
-6
mbar l/s".
F
4
Isolation valve
without
with
code
0
V
Gaz
Helium
code
4
Decades
10
-9
10
-8
10
-7
10
-6
10
-5
10
-4
code
9
8
7
3
6
5
4
2
1
Flange
DN 16
DN 25
DN 40
DN 50
code
0
1
2
3
(1)
Specific value
yes
no
code
1
0
(1) DN 50 not available
for decades 10
-5
and
10
-4
mbar l/s
Used to connect the
calibrated leak and
the sniffer probe for a
calibration.
C2
Calibration accessory
in sniffing
Model
Part No
DN 16
110715
DN 25
110716
adixen Vacuum Products - ASM 380 Operating instructions