Test Menu
Air inlet (Cdt)
Connect an air inlet vent (or nitrogen) at the inlet vent allows to
decrease the He pollution or water pollution of the detector.
The automatic closing after a defined time allows to limit the dry
air or nitrogen consummation if this one is connected.
Always allocate a function key to the inlet vent function to active
it manually.
•
•
•
This function can be harmful when the detector is connected to an ultra
vacuum or semiconductor or thin film chamber! It is important to not do an
inlet vent.
In this case, select "Operator", or / and remove the dedicated function key.
The operator will not have the possibility to do an inlet vent, except with
this menu (which can be locked with password
C 405).
To block the air inlet valve command, delete
Inlet Vent
function key
(
C 405
)
.
The icon
will stay in standard
screen as indicator but the manual activation by operator will be
deactivated.
•