GB 03659 - Edition 02 - March 12
11/13
B 240
Accessories installation
G = Ground
Detectors with
P0344 board
(1)
Detectors with
P0419 board
Internal ground
15
17 - 18
External ground
21 - 24
19 - 36 - 37
Inputs (orders)
Interface
23
Contact open: the filament command is valid.
Contact closed: the filament command is invalid.
Autocalibration sequence start
22
11
Cycle start
8
30
Filament on
20
12
Gross leak mode selection
18
31
Sniffing mode selection
17
13
Vent mode selection
16
32
Outputs (signals)
Sniffer mode (LDS)
1 - 2
5 - 24
Gross leak mode
3 - 4
4 - 23
High sensitivity mode
5 - 6
3 - 22
Test cycle start
7 - 9
2 - 21
Filament on
10 - 11
9 - 28
Helium signal > Reject Set point
12 - 13
1 - 20
Analog output 0/10 V DC (inlet pressure)
19 - 15
0/10 V
0/8 V DC analog output (Helium signal)
14 - 15
0/8 V
36 - G
0/10 V
(2) ASM 182/192 series – ASM 1002
adixen Vacuum Products - ASM 380 Operating instructions