GB 03341 - Edition 03 - March 2010
1/5
C 402
Test Menu
Adjustment of the next parameters:
method and test mode
correction factor
cycle end
air inlet
Purpose
2 test methods proposed:
hard vacuum test
sniffing test
A 200
Make sure that the tested parts can withstand the pressure
difference due to the hard vacuum test.
Test methods
Hard vacuum test
The correction factor allows correction of the measured signal by the
leak detector when it is in parallel with a pump.
Activate the correction factor function:
or
Set the correction factor to apply.
Correction factor
Procedure
Only the digital display is corrected by the correction factor.
The correction factor does not apply to the bargraph display.
Digital and bargraph
display
Quick access to the method choice from the control panel:
•
adixen Vacuum Products - ASM 380 Operating instructions