8/9
B 110
Unpacking/Packaging - Storage -
Handling
GB 04266 - Edition 03 - March 12
3 straps are necessary to close the
box.
Refer to
B 400
.
DN 40 KF hose holder (delivered alone without
centering ring, o'ring and clamp) allows to
fix the hose connecting the inlet detector to
the installation to be tested, during a moving
or an extended storage and to keep under
vacuum inlet detector and hose.
The holder is fixed instead of a lifting ring: unscrew one lifting
ring and screw the holder at its place.
Contact your Customer Center for more informations.
Before any starting up or after an extended storage above 6
months, we advise to change all o'rings (contact your Customer
Center).
Refer to the switching off recommended procedure:
B 400
.
Remote control fixing
Hose holder fixing
Storage
For prolonged storage, factors such as humidity, temperature, saline
atmosphere, etc.. may damage the detector elements.
Seal kits must be stored away from heat and light (direct sunlight
ultraviolet radiation) in order to prevent any elastomers setting.
Packaging for
shipment (Cdt)
adixen Vacuum Products - ASM 380 Operating instructions