GB 04338 - Edition 03 - March 12
5/17
C 406
Advanced Menu
Calibration checking
If calibration = "operator"/"starting", the function "calibration
checking" does a calibration checking according to set parameters.
The calibration checking is deactivated if calibration ="manual".
The calibration checking is done with the internal calibrated leak
(location parameter = internal).
The calibration checking function allows to the operator to gain
time when a calibration is requested but not necessary because the
calibration checking is faster than the complete calibration.
The leak detector compares the internal calibrated leak delivered
signal to the calibrated leak parameters memorized:
If the value is correct, the leak detector calibration is good.
If the value is out of limits, there is a message which request to
start a complete leak detector calibration.
Function activation.
Set points launching the
calibration checking. The
first reached will launch the
checking.
s¬
At any time, the operator can launch a leak detector calibration
checking: detector in stand-by mode, press twice on the function
key
in less than 5 secondes.
adixen Vacuum Products - ASM 380 Operating instructions