C 403
Spectro Menu
GB 03317 - Edition 05 - May 12
Calibrated leak
It is necessary to set the calibrated
leak used for the autocalibration:
use the information indicated on the
calibrated leak label or its calibration
certificate.
In case of leak change, it is necessary
to update these parameters.
According to the internal or
external location selected and
tracer gas selected, the parameters
displayed correspond to the
parameters of the correspondent
leak memorized.
location = internal –> the
autocalibration will be done with
the internal calibrated leak of the
leak detector (He leak only).
location = external –> the autocalibration will be done with an
external calibrated leak (leak He,
3
He or Hy).
The parameter "Calibration valve" allows to open/close the
calibration for a manual calibration for example : this type of
calibration is kept to experts. Do not forget to close it back this
valve after operation.
When a parameters saving is realized (
C 406
), all data of all
calibrated leaks set (1 internal and 3 external) are saved.
More informations about the calibrated leaks
E 412.
adixen Vacuum Products - ASM 380 Operating instructions