8/17
C 406
Advanced Menu
GB 04338 - Edition 03 - March 12
This function allows to calibrate the detector internal gauge.
Leak detection:
Internal Pirani gauge
calibration
Setting
"Cycle end" function = operator
(
C 402
).
Close the detector inlet with a
blank-off flange. Be sure that the
detector is in hard vacuum test and
high sensitiviy mode( (
C 404
).
Limit pressure adjustment
Make sure that the internal
pressure is widely lower than
10
-3
mbar.
Start a cycle: press
The value "Pressure" decreases:
wait the stabilisation of this value
(around 5 minutes) and press on
the key
.
Stop the cycle: press
Make sure that the detector is at
atmospheric pressure.
The value "Pressure" increases:
wait the stabilisation of this value
(around 5 minutes) and press on
the key
.