6.3
Bias Calibration
Bias Calibration is used to offset the deviation between the output measurement and
the output setting. Follow the steps below to perform Bias Calibration:
1.
Ensure the product output is stopped.
2.
Go to
Setting
→
Calibration
and press
to enter the All Bias page.
Figure 6.21 All Bias page
3.
Press
Update
, and the system will generate the
New Value
.
Figure 6.22 Generates the New Value