![Pic Solution helpRAPID Instructions For Use Manual Download Page 2](http://html1.mh-extra.com/html/pic-solution/helprapid/helprapid_instructions-for-use-manual_1554853002.webp)
Guidance and manufacturer’s declaration - electromagnetic immunity
The helpRAPID (
02022527000100) is intended for use in the electromagnetic environment specified below. The
customer or the user of the helpRAPID (
02022527000100) should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test level
Compliance level
Electromagnetic environment -
guidance
Conducted RF
IEC 61000-4-6
150 kHz to 80 MHz:
3Vrms
6Vrms (in ISM and
amateur radio bands)
80% Am at 1kHz
150 kHz to 80 MHz:
3Vrms
6Vrms (in ISM and
amateur radio bands)
80% Am at 1kHz
Portable and mobile RF communications
equipment should be used no closer to
any part of the device, including cables,
than the recommended separation distan-
ce calculated from the equation appropria-
te for the frequency of the transmitter.
Recommended separation distances:
d = 0.35 √P
d = 1.2 √P
Radiated RF
IEC 61000-4-3
10V/m,
80% Am at 1kHz
10V/m,
80% Am at 1kHz
80 MHz to 800 MHz
d = 1.2 √P
800 MHz to 2,7 GHz
d = 2.3 √P
Where p is the maximum output power
rating of the transmitter in watts (W) accor-
ding to the transmitter manufacturer and
d is the recommended separation distance
in metres (m). Field strengths from fixed RF
transmitters, as determined by an electro-
magnetic site survey,
a
should be less than
the compliance level in each frequency
range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular / cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To
assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be consi-
dered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance
level above, the device should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the device.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications equipment and the device.
The helpRAPID (
02022527000100) is intended for use in an electromagnetic environment in which radiated RF distur-
bances are controlled. The customer or the user of the helpRAPID (
02022527000100) can help prevent electroma-
gnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment
(transmittters) and the helpRAPID (
02022527000100) as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum output
power of transmitter
(W)
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d=3.5
√P
80MHz to 800MHz
d=1.2
√P
800MHz to 2.7GHz
d=2.3
√P
0,01
0.12
0.12
0.23
0,1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Guidance and manufacturer’s declaration - electromagnetic immunity
The
helpRAPID (
02022527000100)
is intended for use in the electromagnetic environment specified below. The
customer or the user of the
helpRAPID (
02022527000100)
, should assure that it is used in such an environment.
Radiated RF
IEC61000-4-3
(Test
specifications
for
ENCLOSURE
PORT
IMMUNITY to
RF wireless
communica-
tions
equipment)
Test frequency
(MHz)
Band
a
)
(MHz)
Service
a
)
Modulation
b
)
Modulation
b
) (W)
Distance (m)
IMMUNITY
TEST LEVEL
(V/m)
385
380 – 390
TETRA 400
Pulse
modulation
b
)
18 Hz
1,8
0,3
27
450
430 – 470
GMRS 460,
FRS 460
FM c)
±5 kHz
deviation
1 kHz sine
2
0,3
28
710
704 – 787
LTE Band
13, 17
Pulse
modulation
b
)
217 Hz
0,2
0,3
9
745
780
810
800 – 960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation
b
)
18 Hz
2
0,3
28
870
930
1 720
1 700 –
1 990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25; UMTS
Pulse
modulation
b
)
217 Hz
2
0,3
28
1 845
1 970
2 450
2 400 –
2 570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
b
)
217 Hz
2
0,3
28
5 240
5 100 –
5 800
WLAN 802.11
a/n
Pulse
modulation
b
)
217 Hz
0,2
0,3
9
5 500
5 785
NOTE: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the
ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it does not
represent actual modulation, it would be worst case.
The MANUFACTURER should consider reducing the minimum separation distance, based on RISK MANAGEMENT,
and using higher IMMUNITY TEST LEVELS that are appropriate for the reduced minimum separation distance. Mi-
nimum separation distances for higher IMMUNITY TEST LEVELS shall be calculated using the following equation:
E = √P
6
d
Where P is the maximum power in W, d is the minimum separation distance in m, and E is the IMMUNITY TEST
LEVEL in V/m.
Safety Information
The symbols below can be found in the Instruction Manual as well as on labels and other components.
“PLEASE REFER TO THE INSTRUCTIONS
FOR USE” symbol
“TYPE BF APPLIED PART” symbol
0068
Symbol for “COMPLIES WITH MDD
93/42/EEC as amended by directive
2007/47/EC”
Storage and transport humidity
symbol
“MANUFACTURER” symbol
Storage and transport temperature
symbol
SN
Symbol for “SERIAL NUMBER”
Caution
Symbol for “DIRECT CURRENT”
Lot number
Latex free
Catalogue Number
IP21
Protected against solid foreing objects
of 12.5mm Ø and greater and protected
against vertically falling water drops.
Atmospheric pressure
Appliance date of production: the second pair of digits in the lot number identifies the year, the third pair identifies the month.
(
1016100001; 16 = 2016; 10 = October)
THE PRODUCT COMPLIES WITH:
EN 1060-1: 1995/A2: 2009 Non-invasive sphygmomanometers – Part 1 General requirements
EN 1060-3: 1997/A2:2009 Non-invasive sphygmomanometers – Part 3: Supplementary requirements for electro-mechanical
blood pressure measuring systems.
EN 1060-4: 2004 Non-invasive sphygmomanometers – Part 4: Procedures to determine the overall system accuracy of
automated non-invasive sphygmomanometers.
THIS PRODUCT COMPLIES WITH THE DIRECTIVE 2012/19/EU.
The crossed bin symbol on the appliance indicates that the product, at the end of its life, must be disposed of
separately from domestic waste, either by taking it to a separate waste disposal site for electric and electronic ap-
pliances or by returning it to your dealer when you buy another similar appliance. The user is responsible for taking
the appliance to a special waste disposal site at the end of its life. If the disused appliance is collected correctly
as separate waste, it can be recycled, treated and disposed of ecologically; this avoids a negative impact on both
the environment and health, and contributes towards the recycling of the product’s materials. For further information regarding
the waste disposal services available, contact your local waste disposal agency or the shop where you bought the appliance.
THIS PRODUCT COMPLIES WITH EU DIRECTIVE 2006/66/EC
The crossed bin symbol on the batteries indicates that, at the end of their life, they must be disposed of sepa-
rately from domestic waste, either by taking them to a separate waste disposal site for batteries or by returning
them to your dealer when you buy similar rechargeable or non-rechargeable batteries. The chemical symbols
Hg, Cd, Pb, printed under the crossed bin symbol, indicate the type of substance contained in the batteries:
Hg=Mercury, Cd=Cadmium, Pb=Lead. The user is responsible for bringing batteries, at the end of their life, to the appropriate
collection facilities in order to facilitate treatment and recycling. The correct disposal of your old toy will help prevent potential
negative consequences for the environment and human health and favour the recycling of the materials of which the product
is made. Illegal disposal of the product by the user will damage the environment and human health. For further information
regarding the waste disposal services available, contact your local waste disposal agency or the shop where you bought the
appliance.
76 022527 000 100
Rev
. 01/2019
02 022527 000 100
ينصلا يف عنص
0068
PIKDARE S.p.A.
Via Saldarini Catelli, 10
22070 - Casnate con Bernate (CO) - Italy
www.picsolution.com
Pouch cleaning symbols
Wash by hand in cold water,
30° C
Do not bleach,
Do not tumble dry,
Line drying,
Do not iron,