Pic Solution helpRAPID Instructions For Use Manual Download Page 1

Soft Cuff

Fig. D

Fig. E

Automatic Blood pressure monitor

THESE INSTRUCTIONS ARE IMPORTANT. PLEASE KEEP THEM FOR FUTURE REFERENCE.

Dear customer, 
Thank you for choosing Pic Solutions helpRAPID, an automatic blood pressure monitor designed and manufactured according 
the most up-to-date technologies. Before using the device, we highly recommend that you read the short list of warnings 
contained in this manual, to make sure you have correctly understood how to use it.

INTRODUCTION

The regular use of a Blood Pressure Monitor allows you to monitor blood pressure continuously, but it should not take the place of 
regular check-ups by your doctor. We recommend that you visit your doctor regularly for general check-ups and for more detailed in-
formation on blood pressure. Blood pressure measurements determined with this unit are equivalent to those obtained by a trained 
observer using cuff/ stethoscope auscultation method, within the limits prescribed by the international ESH Validation protocol.

WARRANTY 

The product is guaranteed for 5 years against all conformity defects under the normal utilisation conditions provided in the user instruc-
tions. Therefore, the warranty shall not apply in case of damages caused by improper use, wear, or accidental events. Batteries, cuff and 
wearing parts are not included in the warranty. In case of malfunction, contact exclusively the authorised Pic Technical 

Customer Service.

RAPID-TECH Comfort Experience

TM

This unit uses the oscillometric method to detect your blood pressure.  

RAPID-TECH Comfort Experience

TM

 is the latest generation 

technology chosen by Pic for the blood pressure monitors of the line RAPID. It allows fast and, therefore, less annoying monitoring 
because the measurement is performed already during the armband inflation phase. Pic helpRAPID has been clinically validated in 
accordance with the  ESH (European Society of Hypertension) protocol, as a guarantee of its accuracy and precision. 

NAME/FUNCTION OF EACH PART

List of included items: Monitor, Pic Ready Cuff M-L size (

 02007832000000), 4xAAA batteries, User manuals, Carring bag.

Features:
• 128x50mm Blue LCD display with White backlight
• Touch sensor keys
• Two memory areas
• 100 recorded measurement per area
• Inflation-type measurement 

What is blood pressure? 

Blood pressure is the pressure exerted on the artery tube while blood flows through the arteries. The pressure measured when 
the heart contracts and sends blood out of the heart is systolic (highest) blood pressure. The pressure measured when the 
heart dilates with blood flowing back into the heart is called diastolic (lowest) blood pressure. 

Why measure your blood pressure?

Among the various health problems afflicting modern people, problems associated with high blood pressure are by far the 
most common. High blood pressure’s dangerously strong correlation with cardiovascular diseases and high morbidity has 
made measuring blood pressures a necessity of identifying those at risk. 

Blood Pressure Fluctuation 

Blood pressure fluctuates all the time! 
You should not be over worried if you encountered two or three measurements at high levels. Blood pressure changes over 
the month and even throughout the day. It is also influenced by the season, temperature and activities. 

Classification of high blood pressure

The out-of-office blood pressure classification published by ESH/ESC in 2013 is as follows:

Note:

 It is not intended to provide a basis of any type of rush toward emergency conditions/diagnosis based on the scheme: 

the scheme is meant only to discriminate between the different levels of blood pressure.

WARNING 

Only a physician can tell your normal BP range. Please 
contact a physician if your measuring result falls out of 

the range. Kindly note that only a physician could tell whether 
your blood pressure value has reached a dangerous point.

 

Irregular Heartbeat Detector

This blood pressure monitor is equipped with an algorithm for 
Irregular Heartbeat (IHB) Detection. An irregular heartbeat is de-
tected when a heartbeat rhythm varies while the unit is measur-
ing the systolic and diastolic blood pressure. During each meas-
urement, the monitor records all the pulse intervals and calculate 
the average; if there are two or more pulse intervals, the difference 
between each interval and the average is more than the average 

value of ±25%, or there are four or more pulse inter vals, the difference between each interval and the average is more than 
the average value of ±15%, then the irregular heartbeat symbol will appear on the display with the measurement result. 

WARNING  

The appearance of the IHB icon indicates that a pulse irregularity consistent with an irregular heartbeat was detected 
during measurement. Usually this is NOT a cause for concern. However, if the symbol appears often, we recommend 

you seek medical advice. We wish to underline that the device does not replace a visit to the cardiologist, but it is useful to 
detect any irregularities in advance.

Why the blood pressure I get from the hospital is different from home?

The blood pressure is different even during 24 hour because of the weather, emotion, exercise etc, specially the “white coat” in 
hospital which makes the results are higher than the ones at home. 

Is the result the same if measuring on the right arm?

Measurements can be taken on either arm; results can however differ sometimes so it is always advisable to use the same arm.

LCD display explanation

At starting check the completeness of the LCD segments.

SYMBOL

DESCRIPTION

EXPLANATION

Systolic pressure

Maximum/systolic pressure result

Diastolic pressure

Minimum/diastolic pressure result

Pulse

Pulse/minute

Deflating

The arm cuff is deflating

Clock (hours: minutes)

Currently time

Memory

If “M” shows that displayed measurement values is from the memory.

mmHg

Blood pressure unit of measurement mmHg

Low battery

Batteries are low and need to be replaced

Motion detected while measuring Shocking will result in inaccurate

Average

The average of last 3 measurements taken within 30 minutes.

Recalling

The records will be showed

Irregular heartbeat

Irregular heartbeat

Memory area A

The Area A is chosen

Memory area B

The Area B is chosen

Grade

Blood pressure level indicator

Date

“M” shows the month, “D” shows the day

Heartbeat

Heartbeat detection during the measurement

STAR

T

STOP

SET

MEM

A

A

No

rm

al

Pul

/min

mmHg

DIA

mmHg

SYS

B

The Choice of Power Supply

1. Battery powered mode: 4 batteries 1,5V AAA; 6V

 

   

2.  Power supply via mains adapter: Input: 100 - 240V~, 50 / 60Hz,  0.2A max, Output: 6V 

 1A

 Can be powered only with the “Adapter for blood pressure monitor” 

 

02000520000000 

(BLJ06L060100P-V) with European plug and the “Adapter for blood pressure monitor UK” 

 

 

02010310000000 (BLJ06L060100P-B) with UK plug (Not included).
Please unplug the adapter to depart from the using utility power.
 NB: The adapter connector is located on the right side of the monitor. Place NO obstacles on the 
right side for easy pull-off adapter.

WARNING

In order to protect your device and yourself and to get the best performances, use only the batteries and adapter 

according to the information provided in this manual (4 batteries 1,5V AAA; 6 V DC).

 Installing and Replacing the Batteries

 1. Slide off the battery cover.

 2. Install the batteries by matching the correct polarity, as shown.

 3. Replace the cover.

Replace the batteries whenever the below happen

Is displayed 

, the display dims, the display does not light up.

WARNING

• If the device is not used for a while, please remove the batteries.

•  The old battery is harmful to the environment, so please disposal accordingly to the indication reported in this user 

manual (EU Directive 2006/66/EC).

• Do not dispose of batteries in fire. Batteries may explode or leak

• For instructions on battery disposal please refer to the indication reported in this user manual.

• There are no user serviceable parts inside. Batteries or damage from old batteries is not covered by warranty.

•  Only use brand batteries. Always replace with new batteries together. Use batteries of the same brand and same type. Other 

types of batteries can cause explosions or fires.

•  The battery can explode if handled carelessly. Never recharge, remove or position the batteries near sources of heat. Do not incinerate.

• Rechargeable batteries are not recommended.

Setting Date and Time

It is important to set the clock before using your blood pressure monitor, so that a time stamp can be assigned to each meas-

urement that is stored in the memory. (year: 2015-2055, time: 12H/24H). In the setting mode, use the “MEM” button to change 

the value and the “SET” button to confirm your choice.

1. When the unit is off, hold pressing “SET” for 3 seconds to enter the mode for year setting.

2. Press the “MEM” to change the [YEAR].

3. When you get the right year, press “SET” to set down and turn to next step.

4. Repeat step 2 and 3 to set the [MONTH] and [DAY].

5.  Then it will enter time setting mode. There are two kinds of time format “12” hours and “24 “ hours. Press ”MEM” button to 

change the time format. Press “SET” button to confirm the time format.

6. Repeat step 2 and 3 to set the [HOUR] and [MINUTE].

7. After confirming the time, the LCD will display “dOnE” and the monitor will shut off.

Wear the cuff

Plug in the cuff connecting tube into the unit.  Wrap the arm cuff and follow these steps:

1. Use your  right hand to open the cuff  (Fig. A).

2. Place the cuff on your left arm by using the red band as a reference (Fig. B).

 

             

3.  Turn the palm of the hand upward and rest the edge of the cuff at about 2-3 cm (0.8” - 1.2”) over the hollow of the elbow. 

Tighten the cuff (Fig. C).

4.  Position the artery mark   over the main artery (on the inside of your arm). Note: Locate the main artery by pressing with 

2 fingers approximately 2 cm above the bend of your elbow on the inside of your left arm. Identify where the pulse can be 

felt the strongest. This is your main artery. 

5.  Lay your arm on a table, so the cuff is at the same height as your heart. Make sure the tube is not kinked and the edge of the cuff is not folded.
Note: If you use the soft cuff 

 02007827000000 not included, please refer to Fig. D, E and C, instead of Fig. A, B and C.

Helpful Tips

Here are a few helpful tips to help you obtain more accurate readings:

• Do not measure your blood pressure immediately after consuming a large meal.

• To obtain more accurate readings, please wait one hour after meal before measuring.

• Do not smoke or drink tea, coffee or alcohol before measuring your blood pressure.

• Do not measure your blood pressure immediately after taking a bath: wait at least 20 minutes.

• You should not be physically tired or exhausted while taking a measurement.

• Do not measure your blood pressure when you want to discharge urine.

• It is important that you relax during measurement.

• Do not take measurements if you are under stress or tense.

•  Take your blood pressure at normal body temperature. If you are feeling cold or hot, wait a while before taking a measurement.

• Rest for 5 minutes before measuring.

• Wait at least 3 minutes between measurements. This allows your blood circulation to recover.

• Take the measurement in a silent room.

• Do not move and do not talk when taking a measurement

• Do not cross your legs and keep your feet on the ground.

• Keep your back against the backrest of the chair.

•  For a meaningful comparison, try to measure under similar conditions. For example, take daily measurements at approxi-

mately the same time, on the same arm, or as directed by a physician.

   

Start the Measurement

1.

  When the monitor is off, select User A or User B, press the 

 

to turn on the monitor and start the measurement.

LCD display.

Zero setting.

Inflating and measuring.

Display and save the results.

2. Press the 

 to power off, otherwise it will turn off within 1 minute.

If  the 

 icon appears, please, read carefully the Helpful Tips section and repeat the measurement.

Notes:

 To stop the measurement at any time, press the START/STOP button. The cuff will quickly deflate 

Store the cuff carefully.

Recall the measurements

1.

  When the Blood pressure monitor is off, select area A or area B and 

press the “MEM” button to see the average of the last 3 recordings 
taken withn 30 minutes. If the recorded measurements are less than 
3 as well as the last 3 recordings are taken in more than 30 minutes, 
you will see the latest reading.

The symbol of “AVG” will show in the right corner

2. 

Press “MEM” or “SET” to see the recording wanted.

Date and time will display alternately.

 

WARNING

The most recent recording (0) is shown first. Every time a new 
recording is made, the first (0) record is assigned to it. All the 

others are moved by one figure (e.g. 2 becomes 3, and so on), whereas 
the last record (99/99) is deleted from the list.

Delete the measurements

If you did not get the correct measurement, you can delete all results by following below steps
1.  When the blood pressure monitor is in memory mode, select area A or area B and keep the “MEM” button pressed for 3 seconds; 

the display blinks. Press “SET” to confirm deleting and the monitor will turn off.

2. If you don’t want to delete the records, press “START/STOP” to escape
3. If there is no record the display will show “- -“.

Maintenance

In order to get the best performance from your monitor, please follow the below instructions.
• Put in a dry place and avoid the sunshine.
• Avoid touching water, clean it with a dry cloth in case.
• Avoid the intense shaking and collision.
• Avoid the dusty and unstable temperature environment.
• Using the wet clothing to remove the dirt.
• Avoid washing the cuff.

Calibration and service

•  The accuracy of this monitor has been carefully tested and is has been designed for a long life. Indeed the monitor can 

maintain the safety and performance characteristics for a minimum of 10,000 measurements or five years of normal use.

•  It is generally recommended  to check the unit every two years to ensure correct functioning and accuracy. Please consult the 

Contact information in this user manual to get in touch with your local authorized Service Center. 

WARNINGS

•  This device is intended for no-invasive measuring and monitoring of arterial blood pressure on adults, in home and 

professional environment.

• This device is not suitable for use on newborns, children or pregnant women. The device is not validated to be used on the 

above listed patient and it could be dangerous. The use on teenagers shall be evaluated by physician.

• The device cannot be used on pre-eclamptic  female.
• The device is not suitable for use if  your arm circumference is not included in the range 22cm-42cm (8¾’’ - 16½’’).
• People with reduced cognitive abilities shall use the device under supervision.
• Keep out of reach of children, people with reduced cognitive abilities or pets: the device contains small parts which can 

cause suffocation or internal lesions if swallowed. The cuff tubes and the cable of AC/DC adapter (optional) can create stran-
gulation risk. 

• If the arm cuff pressure  reaches 300 mmHg it deflate automatically. If the arm cuff does not deflate when pressure  reaches 

300 mmHg or if you feel discomfort during the measurement take it off your arm and press the START/STOP button to stop 
inflation.

• Do not confuse self-monitoring with self-diagnosis. By using this device you can monitor your blood pressure, but it does not 

replace your doctor. Begin or end medical treatment based solely on medical advice. If you are taking medication, consult 
your physician to determine the most appropriate time to measure your blood pressure. Never change a prescribed medi-
cation without consulting your Physician.

• If Irregular Heartbeat (IHB) brought by common arrhythmias is detected in the procedure of blood pressure measurement, 

a signal of will be displayed. Under this condition, the Electronic Sphygmomanometer can keep function, but the results 
may not be accurate, it’s suggested that you consult with your physician for accurate assessment. When the device is used 
for measurements on patients with common arrhythmias, such as atrial or ventricular premature beats or atrial fibrillation, a 
deviation may occur. Please consult your doctor about the result.

• Please consult your physician before using the device if you have active implanted medical devices.
• This unit is not suitable for continuous monitoring during medical emergencies or operations. 
• This device cannot be used with HF surgical equipment at the same time.
• This device is not intended to be used during  patient transportation outside a healthcare facility.
• It is not intended for use on extremities other than   the arm or for functions other than obtaining a blood pressure measurement.
• To avoid measurement errors, please avoid the condition of strong electromagnetic field radiated interference signal or 

electrical fast transient/burst signal.

• The user must check that the equipment functions safely and see that it is in proper working condition before being used.
• Manufacturer will make available on request circuit diagrams, component parts list etc. only to qualified person.
• Too frequent measurements can cause injury to the PATIENT due to blood flow interference. Please check (by observation 

of the limb concerned) that operation of helpRAPID does not result in prolonged impairment of the circulation of the blood 
of the PATIENT.

• This unit is not suitable for continuous monitoring during medical emergencies or operations. Otherwise, the patient’s arm 

and fingers will become anaesthetic, swollen and even purple due to a lack of blood.

• Check that the device tubes are not kinked while using the device: kinked tubes can generate lack of blood in the extremities.
• Please use the device under the environment which was provided in the instruction manual. If this is not complied with, 

correct operation and life of the device may be compromised.

• During using, the patient will contact with the cuff. The materials of the cuff have been tested and found to comply with 

requirements of ISO 10993-5:2009 and ISO 10993-10:2010. It will not cause any potential sensization or irritation reaction.

• The application of the CUFF over a wound, as this can cause further injury
• The application of the CUFF and its pressurization on any limb where intravascular access or therapy, or an arterio-venous 

(A-V) shunt, is present because of temporary interference to blood flow and could result in injury to the PATIENT

• The application of the CUFF and its pressurization on the arm on the side of a mastectomy can cause injury or give not 

accurate measurement.

• Do not inflate the cuff on the same limb which other monitoring ME equipment is applied around simultaneously, because 

this could cause temporary loss of function of the simultaneously-used monitoring ME equipment.

• The device is a non-AP/APG device and not suitable for use in the presence of anaesthetic mixtures that are inflammable 

with air, oxygen or nitrous oxide.

• The operator shall not touch DC output jack of AC/DC adapter and the patient simultaneously.
• Please use accessories and detachable parts specified/authorized by the manufacturer. Otherwise, it may cause damage to 

the unit or danger to the user/patients.

• We recommend checking its performance every two years or after it has been repaired. The device must be checked again 

in case of damage due to impact (such as falls) or exposure to liquids and/or extreme temperatures (hot/cold) or extreme 
humidity variations. 

• For public place use, please contact your local distributor or the Italian toll-free number 800 900 080 to get support how to proceed.
• Please dispose of accessories, detachable parts, and the ME EQUIPMENT according to the local guidelines.
• In case of problems concerning performances, settings, maintenance or use of the device, contact your local distributor or the 

Italian toll-free number 800 900 080. Warning! Do not service and/or repair the appliance while it is being used! Any mainte-
nance/service activity/operation must be done solely by a PiC Service Centre. No modification of the equipment are allowed.

• When the device is stored at the minimum/maximum storage temperature It takes 4h to warm/cool until the device is ready 

for its intended use.

This section includes a list of error messages and frequently asked questions for problems you may encounter with your 

sphygmomanometer. If the products not operating as you think it should, check here before arranging for servicing.

PROBLEM

SYMPTOM

CHECK THIS

REMEDY

No power

Display will not 

light up.

Batteries are exhausted.

Replace with new batteries

Batteries are inserted incorrectly.

Insert the batteries correctly

AC adaptor is inserted incorrectly.

Insert the AC adaptor correctly

Low 

batteries

The display is dim 
or shows 

Batteries are exhausted.

Replace with new batteries

Error 

message

E 1 shows

The cuff is not secure

Refasten the cuff and then measure again.

E 2 shows

The cuff is very tight

Refasten the cuff and then measure again.

E 3 shows

The pressure of the cuff is excess.

Relax for a moment and then measure again.

E10 or E11 shows

The monitor detected motion while 

measuring.

Movement can affect the measurement. Relax 

for a moment and then measure again.

E20 shows

The measurement process does not 

detect the pulse signal.

Loosen the clothing on the arm and then 

measure again

E21 shows

The treatment of the measurement 

failed.

Relax for a moment and then measure again.

EExx, shows on the 

display.

A calibration error occurred.

Retake the measurement. If the problem 

persists, contact the dealer or our customer 

service department for further assistance. 

Refer to the warranty for contact information 

and return instructions.

OUT

Out of range measurement.

Relax for a moment. Refasten the cuff and 

then measure again. If the problem persists, 

contact your Physician.

 

SPECIFICATIONS

•  Power supply:  

Battery powered mode: 4 batteries 1,5V AAA; 6V

 

  (~ 225 measurements)

  

Power supply via mains adapter: Input: 100 - 240V~, 50 / 60Hz, 0,2A max, Output: 6V 

 1A

    

 Can be powered only with the “Adapter for blood pressure monitor” 

 

 02000520000000 

(BLJ06L060100P-V) with European plug and the “Adapter for blood pressure monitor UK”  

 

02010310000000 (BLJ06L060100P-B) with UK plug (Not included).

•  Display: 

Blue LCD with white backlight, V.A.128x50 mm 

•  Measurement mode:

  

Oscillometric inflation -type measurement mode

•  Measurement range:

  

Rated cuff pressure: 0mmHg~299mmHg

  

Measurement pressure: SYS: 60-230mmHg, DIA: 40-130mmHg

   

pulse value: 40–199 beat/minute

•  Accuracy:

 Pressure: ± 3mmHg (5°C to 40°C)

   

Pulse value: ±5% 

•  Normal working 

 condition: 

Temperatures: from 5°C to 40°C Relative humidity≤85%

   

Atmospheric pressure: 86kPa to 106kPa

•  Storage & transportation 

  condition:  

Temperatures: from -5°C to +50°C. Relative humidity ≤85%

   

Atmospheric Pressure: 50-106 kPa

•  Arm circumference:

 

About 22cm~42cm 

•  Net Weight:  

Approx. 270 g (Excluding the batteries) 

•  External dimensions:

  

Approx.100 × 186 × 35mm

•  Accessories:

    

 Not  included:  Adaptor 

 

 02000520000000 with European plug - 

 

 02010310000000 

with UK plug  and soft cuff 

 

 02007827000000 (size M-L)

   

Included: Pic Ready cuff 

 

 02007832000000 (size M-L)  

•  Mode of operation:  

Continuous operation

•  Useful life:

  

10,000 Measurements 

•  Degree of protection:

   

 Type BF applied part (cuff )

•  Protection against 

   ingress of water:

   

 IP  21 

Protected against solid foreing objects of 12.5mm Ø and greater and protected against 

vertically falling water drops.

•  Software version:

    

A01 

Contact Information

For further information on our products, please visit the website www.picsolution.com. Here you can find customer service, 
solve ordinary problems.

EMC Guidance

1)  This product needs special precautions regarding EMC and needs to be installed and put into service according to the EMC 

information provided, and this unit can be affected by portable and mobile RF communications equipment.

2)  *Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in incorrect 

operation of the unit.

3)  Caution: This unit has been thoroughly tested and inspected to assure proper performance and operation!
4)  * Caution: This machine should not be used adjacent to or stacked with other equipment and that if adjacent or stacked 

use is necessary, this machine should be observed to verify normal operation in the configuration in which it will be used.

Guidance and manufacturer’s declaration – electromagnetic emissions

The helpRAPID (

 02022527000100) is intended for use in the electromagnetic environment specified below. The 

customer or the user of the helpRAPID (

 02022527000100) should assure that it is used in such an environment.

Emissions test

Compliance

Electromagnetic environment - guidance

RF emissions

CISPR 11

Group 1

The device uses RF energy only for its internal function. 

Therefore, its RF emissions are very low and are not 

likely to cause any interference in nearby electronic 

equipment.

RF emissions

CISPR 11

Class [B]

The  helpRAPID (

  02022527000100) is suitable for 

use in all establishments, other than domestic and tho-

se directly connected to the public low-voltage power 

supply network that supplies buildings used for dome-

stic purposes.

Harmonic emissions

IEC 61000-3-2

Class A

Voltage fluctuations/

flicker emissions

IEC 61000-3-3

Complies

Guidance and manufacturer’s declaration - electromagnetic Immunity

The helpRAPID (

 02022527000100) is intended for use in the electromagnetic environment specified below. The 

customer or the user of the helpRAPID (

 02022527000100) should assure that it is used in such an environment.

Immunity Test

IEC 60601 Test level

Compliance level

Electromagnetic environment - 

guidance

Electrostatic

discharge (ESD)

IEC 61000-4-2

±8 kV contact

±15 kV air

±8 kV contact

±15 kV air

Floors should be wood, concrete or cera-

mic tile. If floors are covered with synthetic 

material, the relative humidity should be at 

least 30%.

Electrical fast

transient/burst

IEC 61000-4-4

Power supply lines: ±2 kV

Input/output lines: ±1 kV

Power supply lines:

±2 kV

Mains power quality should be that of a 

typical commercial or hospital environ-

ment.

Surge

IEC 61000-4-5

line(s) to line(s): ±1 kV

line(s) to earth: ±2 kV

100 kHz repetition 

frequency

line(s) to line(s): ±1 kV

100 kHz repetition

frequency

Mains power quality should be that of a 

typical commercial or hospital environ-

ment.

Voltage dips, short 

interruptions and 

voltage variations on 

power supply input 

lines IEC 61000-4-11

0%U

T

; 0.5 cycle

At 0°, 45°, 90°, 135°,

180°,225°,270° and 315°

0%U

T

 ; 1 cycle

and

70%U

T

 ; 25/30 cycles

Single phase: at 0°

0% U

T

 ; 300 cycle

0% U

T

; 0.5 cycle

At 0°, 45°, 90°, 135°,

180°,225°,270° and 315°

0% U

T

; 1 cycle

and

70% U

T

; 25/30 cycles

Single phase: at 0°

0% U

T

; 300 cycle

Mains power quality should be that of a 

typical commercial or hospital environ-

ment.

Power frequency

(50/60Hz) magnetic 

field IEC 61000-4-8

30 A/m

50Hz/60Hz

30 A/m

50Hz/60Hz

Power frequency magnetic fields should 

be at levels characteristic of a typical lo-

cation in a typical commercial or hospital 

environment.

NOTE U

T

 is the a.c. mains voltage prior to application of the test level.

START

STOP

SET

MEM

A

A

Pul

/min

mmHg

DIA

mmHg

SYS

B

Backlighted LCD 

Mains adapter 
connector  

BLOOD PRESSURE 
GRADE INDICATOR:

Cuff connector

Cuff 

Memory Key 

User selection switch 

START/STOP Key 

Setting Key 

help

RAPID

M100% + Y100%

P 485 C

P 281 C

The current memory 

record is No. 2. Eight 

records in total.

The corresponding 

date is 11th October

The corresponding 

time is 10:38.

INSTRUCTIONS FOR USE
PLEASE REFER TO THE INSTRUCTIONS FOR USE

Normal

High

Hypertension

Normal Blood Pressure

Pic Ready Cuff

Fig. A

Fig. B

Main artery

Cuff tube

2-3cm (0.8”-1.2”)

Fig. C

Summary of Contents for helpRAPID

Page 1: ...or a long life Indeed the monitor can maintain the safety and performance characteristics for a minimum of 10 000 measurements or five years of normal use It is generally recommended to check the unit...

Page 2: ...z 2 0 3 28 870 930 1 720 1 700 1 990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Band 1 3 4 25 UMTS Pulse modulation b 217 Hz 2 0 3 28 1 845 1 970 2 450 2 400 2 570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE...

Page 3: ...liser un chiffon humide pour enlever la salet viter de laver le brassard l eau Calibrage et assistance Cet appareil a t con u pour durer longtemps et sa pr cision a t attentivement test e C est pourqu...

Page 4: ...tion b 18 Hz 2 0 3 28 870 930 1 720 1 700 1 990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE Bande 1 3 4 25 UMTS Pulsations Modulation b 217 Hz 2 0 3 28 1 845 1 970 2 450 2 400 2 570 Bluetooth WLAN 802 11 b g...

Page 5: ...gwa townych wstrz s w i uderze Unika zapylenia i otoczenia o zmiennej temperaturze Zabrudzenia czy ci wilgotnymi szmatkami Unika mycia mankietu Kalibracja i pomoc techniczna Ten ci nieniomierz zosta...

Page 6: ...00100 powinien dopilnowa aby urz dzenie by o u ywane w takim rodowisku Fale radiowe IEC61000 4 3 Badawcze Parametry urz dzenia dla PORTU OBUDOWY ODPORNO na bezprze wodow radiow komunikacj sprz tu Cz s...

Page 7: ...T STOP 3 10 000 22cm 42cm 8 16 AC DC 300 mmHg 300 mmHg START STOP IHB helpRAPID ISO 10993 5 2009 ISO 10993 10 2010 V ME ME AP APG DC AC DC 800 900 080 080 ME 800 900 080 PiC 4 AC AC E 1 E 2 E 3 E10 E1...

Page 8: ...12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 d m P Watt W 1 80 MHz 800 MHz 2 helpRAPID 02022527000100 helpRAPID 02022527000100 RF IEC61000 4 3 MHz Band a MHz Service a b b W m...

Page 9: ...yabilir Genel olarak do ru i levselli i ve hassasiyeti sa lamak i in cihaz n iki y lda bir kontrol edilmesi tavsiye edilir L tfen yerel yetkili Servis Merkezinizle ileti im kurmak i in bu kullan m k l...

Page 10: ...DECT LTE Band 1 3 4 25 UMTS Nab z mod lasyonb 217 Hz 2 0 3 28 1 845 1 970 2 450 2 400 2 570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE Band 7 Nab z mod lasyonb 217 Hz 2 0 3 28 5 240 5 100 5 800 WLAN 8...

Page 11: ...izven obmo ja 22 42 cm 8 16 Osebe z zmanj animi kognitivnimi sposobnostmi morajo napravo uporabljati pod nadzorom Napravo shranjujte zunaj dosega otrok oseb z omejenimi kognitivnimi sposobnostmi in hi...

Page 12: ...e je to potrebno za izpolnitev RAVNI PREIZKUSA ODPORNO STI Testna razdalja 1 m je dovoljena v skladu s standardom IEC 61000 4 3 a Pri nekaterih storitvah so vklju ene samo frekvence v smeri proti baz...

Page 13: ...kontakt informacije u ovom korisni kom uputstvu da stupite u kontakt sa lokalnim ovla c enim servisom UPOZORENJA Ovaj ure aj je namenjen za neinvazivno merenje i prac enje arterijskog krvnog pritiska...

Page 14: ...Modulacijab 18 Hz 2 0 3 28 870 930 1 720 1 700 1 990 GSM 1800 CDMA 1900 GSM 1900 DECT LTE opseg 1 3 4 25 UMTS Puls Modulacijab 217 Hz 2 0 3 28 1 845 1 970 2 450 2 400 2 570 Bluetooth WLAN 802 11 b g n...

Page 15: ...l v dett helyen vja a nedvess gt l Sz ks g eset n t r lje le sz raz ruh val Ker lje az er s r zk d st s t seket Ne t rolja poros vagy h m rs klet ingadoz snak kitett helyen A szennyez d seket nedves...

Page 16: ...s A helpRAPID 02022527000100 k sz l ket az al bb meghat rozott elektrom gneses k rnyezetben t rt n haszn latra tervezt k Az gyf lnek vagy a helpRAPID 02022527000100 felhaszn l j nak kell gondoskodnia...

Page 17: ...op lysningerne i denne manual for at kontakte dit autoriserede Service Center ADVARSLER Dette apparat er beregnet til ikke invasiv m ling og monitorering af arterielt blodtryk hos voksne i hjemmet og...

Page 18: ...lation b 217 Hz 2 0 3 28 1 845 1 970 2 450 2 400 2 570 Bluetooth WLAN 802 11 b g n RFID 2450 LTE b nd 7 Puls modulation b 217 Hz 2 0 3 28 5 240 5 100 5 800 WLAN 802 11 ikke relevant Puls modulation b...

Page 19: ...STOP 3 10 000 22 cm 42 cm 8 16 AC DC 300 mmHg 300 mmHg START STOP IHB helpRAPID ISO 10993 5 2009 ISO 10993 10 2010 V AP APG DC AC DC 800 900 080 800 900 080 PiC 4 AC AC E 1 E 2 E 3 E10 E11 E20 E21 EE...

Page 20: ...0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 d m P W 1 80 MHz 800 MHz 2 helpRAPID 02022527000100 helpRAPID 02022527000100 IEC61000 4 3 MHz a MHz a b b W m V m 385...

Page 21: ...B D M S T A R T S T O P S E T M E M A A N o r m a l P u l m i n m m H g D I A m m H g S Y S B 6 1 5 AAA 4 1 6 0 2 50 60 100 240 2 1 BLJ06L060100P V 02000520000000 BLJ06L060100P B 02010310000000 4 6 1...

Page 22: ...helpRAPID 02022527000100 helpRAPID 02022527000100 b b a a IEC61000 4 3 RF 27 0 3 1 8 b 18 TETRA 400 380 390 385 28 0 3 2 FM c 5 sine 1 GMRS 460 FRS 460 430 470 450 9 0 3 0 2 b 217 LTE 13 17 704 787 71...

Reviews: