LIFEPAK 20e Defibrillator/Monitor Operating Instructions
4-19
©2006-2015 Physio-Control, Inc.
4 Press
SYNC
. Confirm the sync LED blinks with each detected QRS complex.
Note:
Press
SYNC
again to deactivate synchronous mode.
5 Observe the ECG rhythm. Confirm that a triangle sense marker appears near the middle of
each QRS complex. If the sense markers do not appear or are displayed in the wrong
locations (for example, on the T-wave), select another lead. (It is normal for the sense marker
location to vary slightly on each QRS complex.)
6 Prepare the patient’s skin for therapy electrode application as described in Step 3 of
.
7 Connect the therapy electrodes to the therapy cable, and confirm cable connection to the
defibrillator.
8 Apply therapy electrodes to the patient in the anterior-lateral or anterior-posterior position. If
using standard paddles, apply conductive gel to the paddles and place paddles on the
patient’s chest.
9 Press
ENERGY SELECT
.
10 Press
CHARGE
.
Note:
If the
REMOVE TEST PLUG
message appears, disconnect test plug and connect
therapy electrodes to QUIK-COMBO therapy cable and press
CHARGE
.
11 Make certain all personnel, including operator, stand clear of the patient, bed, and any
equipment connected to the patient.
12 Confirm ECG rhythm. Confirm available energy.
13 Press and
hold
the
SHOCK
button on the defibrillator until the
ENERGY DELIVERED
message
appears on the screen. For standard paddles, press and hold both
SHOCK
buttons on the
paddles simultaneously until the
ENERGY DELIVERED
message appears on the screen.
Release buttons. For safety reasons, the
SHOCK
button on the defibrillator is disabled when
using standard paddles.
Note:
To disarm (cancel a charge), press the Speed Dial. The defibrillator disarms
automatically if shock buttons are not pressed within 60 seconds, or if you change the
energy selection after charging begins. Press
CHARGE
to restart charging.
14 Observe patient and ECG rhythm. Repeat procedure from Step 4, if necessary.
WARNING!
Possible lethal arrhythmia.
Ventricular fibrillation may be induced with improper synchronization. DO NOT use the ECG
from another monitor (slaving) to synchronize the LIFEPAK 20e defibrillator/monitor discharge.
Always monitor the patient’s ECG directly through the ECG cable, therapy cable or use the
remote synchronization procedure. Confirm proper placement of the sense markers on the ECG.
Summary of Contents for LIFEPAK 20
Page 2: ...LIFEPAK 20e DEFIBRILLATOR MONITOR Operating Instructions ...
Page 3: ......
Page 4: ...LIFEPAK 20e DEFIBRILLATOR MONITOR OPERATING INSTRUCTIONS ...
Page 15: ......
Page 49: ......
Page 111: ......
Page 155: ......
Page 171: ......
Page 181: ......
Page 183: ......
Page 189: ......
Page 191: ......
Page 195: ......
Page 199: ......
Page 201: ......
Page 205: ......
Page 209: ......
Page 211: ......
Page 213: ......
Page 215: ......
Page 221: ......
Page 226: ......