LIFEPAK 20e Defibrillator/Monitor Operating Instructions
6-13
©2006-2015 Physio-Control, Inc.
Figure 6-3
IrDA Connections
DATA TRANSFER FROM TrueCPR DEVICE
Note:
The TrueCPR device may not be available in all countries. Contact your local
Physio-Control representative for more information.
You can transfer data from the Physio-Control TrueCPR device to Physio-Control post-event
review tools using the TrueCPR device port on the CodeManagement Module. To transfer data:
1 Confirm the LIFEPAK 20e defibrillator/monitor is either turned off or in archives mode.
2 Connect an approved USB cable between the TrueCPR device and the CodeManagement
Module.
Note:
Use only a USB cable that meets the following specifications: USB 2.0 A-Male to
Mini-B 5-pin Male, 28/24 AWB cable with ferrite core (gold plated), 1 m (3 ft).
3 Turn on the TrueCPR device and follow the instructions in the TrueCPR device Instructions
for Use.
Notes:
• It may take up to one minute for the TrueCPR device to display the Data Transfer screen.
• Data transfer takes approximately 3–5 minutes.
• If you attempt to transmit data from the LIFEPAK 20e defibrillator/monitor and the TrueCPR
device simultaneously, the data from the defibrillator will be transmitted first.
IrDA
Adapter/
Computer
Maximum Distance: 1.0 m (3.28 feet)
Defibrillator
Summary of Contents for LIFEPAK 20
Page 2: ...LIFEPAK 20e DEFIBRILLATOR MONITOR Operating Instructions ...
Page 3: ......
Page 4: ...LIFEPAK 20e DEFIBRILLATOR MONITOR OPERATING INSTRUCTIONS ...
Page 15: ......
Page 49: ......
Page 111: ......
Page 155: ......
Page 171: ......
Page 181: ......
Page 183: ......
Page 189: ......
Page 191: ......
Page 195: ......
Page 199: ......
Page 201: ......
Page 205: ......
Page 209: ......
Page 211: ......
Page 213: ......
Page 215: ......
Page 221: ......
Page 226: ......