Appen
dix C
Screen Messages
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
C-5
©2006-2015 Physio-Control, Inc.
STAND CLEAR/PUSH
SHOCK BUTTON
Stand clear of the patient and push the
SHOCK
button.
START CPR
Initiate CPR in AED mode.
SWITCHING PRIMARY
TO LEAD II
PADDLES
lead is not available and you pressed the
ANALYZE
button.
SWITCHING PRIMARY
TO PADDLES
PADDLES
lead is available and you pressed the
ANALYZE
button.
TRANSMISSION
CANCELLED
Data transmission has been cancelled.
TRANSMISSION
COMPLETE
Data transmission completed successfully.
TRANSMISSION
FAILED
Data transmission was not successful.
UNABLE TO TRANSMIT
Unable to send data.
USE ECG LEADS
The device is attempting synchronized cardioversion and
PADDLES
lead was selected.
USER TEST DID NOT
COMPLETE
Test plug not connected to QUIK-COMBO therapy cable, standard
paddles not seated in paddle wells, possible defective therapy cable,
or a problem with the defibrillator during user test.
USER TEST DID NOT
COMPLETE –
CONNECT TO TEST
PLUG
Test plug not connected to QUIK-COMBO therapy cable, standard
paddles not seated in paddle wells, defective therapy cable, or a
problem with the defibrillator during user test.
USER TEST FAILED
An unsuccessful user test has occurred.
USER TEST IN
PROGRESS
Confirms that the user test is in progress.
USER TEST
SUCCEEDED
A successful user test was completed.
XX% TRANSMITTED
Specified percent of the transmission is completed.
Table C-1
Summary of Screen Messages (Continued)
Message
Description
Summary of Contents for LIFEPAK 20
Page 2: ...LIFEPAK 20e DEFIBRILLATOR MONITOR Operating Instructions ...
Page 3: ......
Page 4: ...LIFEPAK 20e DEFIBRILLATOR MONITOR OPERATING INSTRUCTIONS ...
Page 15: ......
Page 49: ......
Page 111: ......
Page 155: ......
Page 171: ......
Page 181: ......
Page 183: ......
Page 189: ......
Page 191: ......
Page 195: ......
Page 199: ......
Page 201: ......
Page 205: ......
Page 209: ......
Page 211: ......
Page 213: ......
Page 215: ......
Page 221: ......
Page 226: ......