Monitoring
3-24
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
15
CO
2
does not appear on
screen when FilterLine is
connected.
CodeManagement Module not
properly connected to
defibrillator.
• Contact qualified service
personnel.
Low battery voltage.
• Connect to AC power.
Defective battery in
CodeManagement Module.
• Contact qualified service
personnel.
CodeManagement Module
malfunction.
• Contact qualified service
personnel.
Table 3-4
Troubleshooting Tips for EtCO
2
(Continued)
Observation
Possible Cause
Corrective Action