LIFEPAK
®
20e Defibrillator/Monitor Operator’s Checklist
Unit Serial No.:________________________
Location: ____________________________
Instruction
Recommended
Corrective Action
Date
Initials
Insert a
in the box after completing each instruction.
1 Check printed result of 3 A.M. daily auto test. (If no self-test results have
printed, refer to Step 10.) If
:
SELF-TEST FAILED
Contact qualified service personnel.
SELF-TEST DID NOT COMPLETE –
PERFORM MANUAL TEST
Connect test plug or secure paddles in well,
and perform manual User Test. If
CONNECT
TO TEST PLUG
message appears, contact
qualified service personnel.
2 Inspect physical condition for:
Foreign substances
Clean the defibrillator.
Damage or cracks
Contact qualified service personnel.
3 Inspect power source for:
AC power connector plugged into unit and
AC power source; AC Mains LEDs on
defibrillator and optional
CodeManagement Module are lit
If AC Mains LEDs are not lit, contact qualified
service personnel.
Broken, loose, or worn power cable
Replace damaged or broken parts.
4 Check therapy and ECG electrodes for:
Use By date
Replace if past expiration date.
Spare electrodes available
Obtain spare electrodes.
5 Examine accessory cables for:
Cracking, damage, broken or bent parts
or pins, and paddle surfaces for pitting
Replace damaged or broken parts.
6 Disconnect the defibrillator from AC power, wait 2 seconds, press ON and check for:
Momentary
SELF-TEST
messages,
illumination of LEDs, and speaker beep
If absent, contact qualified service personnel.
Service LED is lit
If lit, power off device, then power back on.
If service LED stays lit, contact qualified
service personnel.
7 Check ECG printer for:
Adequate paper supply
Replace if necessary.
Ability to print
If not working, contact qualified service
personnel.
8 Confirm therapy cable connected to defibrillator and perform cable check:
If QUIK-COMBO
®
therapy cable is connected:
• Confirm test plug connected to therapy cable.
• Press
ANALYZE
button.
• After
ANALYZING NOW
message, look
for
REMOVE TEST PLUG
message.
If
CONNECT CABLE
or
CONNECT
ELECTRODES
message appears, contact
qualified service personnel.
If HARD PADDLES are connected:
• Confirm paddles properly seated in wells.
• Select 10 J and press the
CHARGE
button.
Note:
Discharging > 10 joules in paddle
wells may damage defibrillator
If
CONNECT CABLE
or
CONNECT
ELECTRODES
message appears, contact
qualified service personnel.
• When fully charged, press the
SHOCK
button on the paddles and look for the
ENERGY DELIVERED
message.
If no energy delivered and shock tone
continues, press Speed Dial to disarm, and
contact qualified service personnel.
9 Reconnect to AC power and then power off the device.
Daily inspection and testing of the defibrillator/monitor using
this Operator’s Checklist is recommended. This form may
be reproduced.
Summary of Contents for LIFEPAK 20
Page 2: ...LIFEPAK 20e DEFIBRILLATOR MONITOR Operating Instructions ...
Page 3: ......
Page 4: ...LIFEPAK 20e DEFIBRILLATOR MONITOR OPERATING INSTRUCTIONS ...
Page 15: ......
Page 49: ......
Page 111: ......
Page 155: ......
Page 171: ......
Page 181: ......
Page 183: ......
Page 189: ......
Page 191: ......
Page 195: ......
Page 199: ......
Page 201: ......
Page 205: ......
Page 209: ......
Page 211: ......
Page 213: ......
Page 215: ......
Page 221: ......
Page 226: ......