Monitoring
3-8
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
Outdated, corroded, or dried-
out electrodes.
• Check date codes on electrode
packages.
• Use only silver/silver chloride
electrodes with Use By dates that
have not passed.
• Leave electrodes in sealed pouch
until time of use.
Loose connection.
Damaged cable or connector/
lead wire.
• Check/reconnect cable
connections.
• Inspect ECG and therapy cables.
• Replace if damaged.
• Check cable with simulator and
replace if malfunction observed.
Misplaced electrodes/lead
wire.
• Confirm correct placement.
• Select lead view with optimal QRS
detection.
Noise because of radio
frequency interference (RFI).
• Check for equipment causing RFI
(such as a radio transmitter) and
relocate or turn off equipment
power.
4
Baseline wander
(low frequency/high
amplitude artifact).
Inadequate skin preparation.
Poor electrode-skin contact.
Diagnostic frequency
response.
• Prepare skin and apply new
electrodes.
• Check electrodes for proper
adhesion.
• Print ECG in monitor frequency
response.
5
Fine baseline artifact
(high frequency/low
amplitude).
Inadequate skin preparation.
Isometric muscle tension in
arms/legs.
• Prepare skin and apply new
electrodes.
• Confirm that limbs are resting on a
supportive surface.
• Check electrodes for proper
adhesion.
6
Systole beeps not
heard or do not occur
with each QRS
complex.
Volume too low.
QRS amplitude too small to
detect.
• Adjust volume.
• Change ECG lead.
7
Monitor displays
dashed lines with no
ECG leads off
messages.
PADDLES
lead selected but
patient connected to ECG
cable.
• Select one of the limb leads.
8
Heart rate (HR)
display different than
pulse rate.
Monitor is detecting the
patient’s internal pacemaker
pulses.
• Prepare skin and apply new
electrodes in different location.
• Select lead view with optimal QRS
detection.
9
Internal pacemaker
pulses difficult to see.
Pulses from pacemaker are
very small. Monitor the visibility
of frequency response limits.
• Turn on internal pacemaker
detector (refer to
• Connect ECG cable and select
ECG lead instead of paddles.
• Print ECG in diagnostic mode
(refer to
Table 3-2
Troubleshooting Tips for ECG Monitoring (Continued)
Observation
Possible Cause
Corrective Action
Summary of Contents for LIFEPAK 20
Page 2: ...LIFEPAK 20e DEFIBRILLATOR MONITOR Operating Instructions ...
Page 3: ......
Page 4: ...LIFEPAK 20e DEFIBRILLATOR MONITOR OPERATING INSTRUCTIONS ...
Page 15: ......
Page 49: ......
Page 111: ......
Page 155: ......
Page 171: ......
Page 181: ......
Page 183: ......
Page 189: ......
Page 191: ......
Page 195: ......
Page 199: ......
Page 201: ......
Page 205: ......
Page 209: ......
Page 211: ......
Page 213: ......
Page 215: ......
Page 221: ......
Page 226: ......