15. Installing the System (A01)/Therapy (A04) PCB
Assembly (continued)
15.12 Connect the W07 - ECG cable from the
parameter bezel to J6 of the system PCB.
15.13 If present, connect the W33 - IP cable from the
parameter bezel to J7 of the system PCB.
15.14 If present, connect the W35 - temperature
cable from the parameter bezel to J7 of the
system PCB.
NOTE:
Route wires: ECG behind the
connector and IP/Temp in front of connector to
reduce possible cable pinch with the front
case.
Connect J6 ECG
cable (W07)
Connect J7 IP
cable (W33)