Title
LIFEPAK 15 Monitor Defibrillator CO2 Upgrade
Page
5 of 33
Doc Type
Service Process Document
Doc ID
3308217
Doc Rev
A
Doc State
Released
State Date
09/08/2011
4.
Disassembling the Case
4.1
If applicable, remove carrying case.
4.2
Remove all cables and patient connections,
and then remove the batteries.
4.3
Remove the four front mounting screws
(3207361-375) from the bumpers. (Torx-15)
4.4
Remove the four rear mounting screws
(3207361-375) from the bumpers. (Torx-15)
4.5
Remove the bumpers (3207307-001 and
3206968-000) from the case and set aside.
Remove Screws
(4 places)
Remove Screws
(4 places)