12. Installing the CO2 (A23) Modules (continued)
12.4 Route the CO2 tubing and CO2 adapter cable
as shown. Ensure that the cable and tubing are
not pinched when installing the bracket cover.
12.5 Attach the bracket cover to the OEM bracket
with four new screws (202253-760), torque to
64 in-oz.
12.6 If Present, route the NIBP cable through the
cable mount on side on bracket cover.