Title
LIFEPAK 15 Monitor Defibrillator CO2 Upgrade
Page
33 of 33
Doc Type
Service Process Document
Doc ID
3308217
Doc Rev
A
Doc State
Released
State Date
09/08/2011
18.5.1 If SpO2 is a monitored channel in any of the monitoring sets, reset the set using the information on printed defaults from section
18.1.
Note:
For more information about monitoring channels setup, refer to LIFEPAK 15 setup options 3306226.
19.
Perform PIP/TCP
Complete Performance Inspection Procedure (PIP) for a LIFEPAK 15 device for all the features that are installed into the
device.
Use LIFEPAK 15 V1/V2 Service Manual, Manual PIP/TCP procedure or CAPIP test
.
20.
Document the Upgrade (
International Service
– document service action per local processes
)
Complete a standard upgrade service report to document all upgrade activities. Perform Post Service Report download.