Title
LIFEPAK 15 Monitor Defibrillator CO2 Upgrade
Page
14 of 33
Doc Type
Service Process Document
Doc ID
3308217
Doc Rev
A
Doc State
Released
State Date
09/08/2011
8.
Removing the OEM Modules bracket assembly
8.1
Remove the three mounting bracket screws.
Discard the screws.
8.2
If present, disconnect the NIBP inlet tube. Lift
the assembly as needed.
8.3
Lift the mounting bracket and cover from the
rear case.
8.4
For the NIBP option disassembly, remove the
NIBP cable (W27) from cable guide on bracket
cover.
8.5
Remove the four screws from the bracket cover
(3206965) and lift off the cover. Discard the
screws.
Remove
Screws
(3 places)
NIBP Tubing
disconnected
Remove NIBP
cable from
cable guide
Remove
screws
(4 places)