13. Installing the OEM PCB (A06)
13.1 Lift the tab at the upper left corner of the OEM
PCB over the ground clip on the power bracket,
and then press down on the tab while sliding
the OEM PCB to the left to engage the direct
connection (J25) to the A03 power PCB at J16.
13.2 Ensure that the ground tab of the OEM PCB is
above the power PCB bracket. Ensure the
OEM PCB sits flat on the OEM bracket and
presses straight into the power PCB.
13.3 Install four new screws (202253-761), torque to
6.8 in-lb.
13.4 Install the nylon standoff (3010569-02) in the
OEM hole (torque to 64 in-oz
using a 1/4”
socket).
13.5 If present, connect the SpO2/OEM cable (W21)
(3207000) and ensure locking feature is
engaged.
13.6 Connect the CO2/OEM cable (W26) (3206998)
and ensure locking feature is engaged.
13.7 If present, connect the NIBP/OEM cable (W27)
(3012181) and lock into place with the retainer
clip (3010805-000).
Align PCB Tab
on top of
metal bracket
Align J25
Connector, slide
toward Power PCB
J16 to connect
Connect SPO2/
OEM Cable (J26)
NIBP Retainer Clip
3010805
Connect NIBP/
OEM Cable (J28)
T
O
R
Q
U
E
T
O
R
Q
U
E
T
O
R
Q
U
E
6.8 In-Lb
T
O
R
Q
U
E
T
O
R
Q
U
E
T
O
R
Q
U
E
64 In-Oz
Nylon Standoff
3010569-02
Screw
202253-761
(4 places)
Connect CO2/
OEM Cable (J27)