Title
LIFEPAK 15 Monitor Defibrillator CO2 Upgrade
Page
9 of 33
Doc Type
Service Process Document
Doc ID
3308217
Doc Rev
A
Doc State
Released
State Date
09/08/2011
5.
Removing the System (A01)/Therapy (A04) PCB
Assembly
5.1
Disconnect the connectors on the system PCB
as follows:
5.1.1 J1 - Press the connector locking tabs and
disconnect the power/system PCB cable
(W01).
5.1.2 (The system/interface PCB cable was
disconnected during case disassembly).
5.1.3 J4 - Disconnect the USB flex connector
(W14).
5.1.4 J6
– Disconnect the ECG connector cable
(W7).
5.1.5 J7 - Disconnect the IP/Temp connector cable
of the IP/Temp wire harness (W33 or W35) (if
present).
5.1.6 J106 (on A22 biphasic module) - Disconnect
the biphasic to therapy cable from the
biphasic PCB (W20).
Disconnect J4
USB Flex
Connector (W14)
Disconnect J6 ECG
and J7 IP/ Temp (if
present) connectors.
(W7 & W33)
Disconnect J1
PWR/ SYS Cable
(W01)
Disconnect J106
BTE cable
connector (W20)