Title
LIFEPAK 15 Monitor Defibrillator CO2 Upgrade
Page
10 of 33
Doc Type
Service Process Document
Doc ID
3308217
Doc Rev
A
Doc State
Released
State Date
09/08/2011
5.
Removing the System (A01)/Therapy (A04) PCB
Assembly (continued)
5.2
Remove the seven screws (202253-761) and
one screw with washer (3207337-312) that
secured the system PCB to the rear case.
Discard the screws.
NOTE:
The screw that is in the ECG shield
can easily be missed as part of screw
removal. Ensure that all screws are removed
prior to lifting the system/therapy PCB
assembly.
5.3
Set the rear case upright and move the
system/therapy PCB assembly towards the
front of the case to gain access to the rear
therapy PCB connectors.
5.4
Disconnect the J22 transfer relay from the
therapy board.
5.5
Disconnect the J21 transfer relay from the
therapy board.
Disconnect
J22 Red &
Black Cable
(A13)
Disconnect J21
Transfer Relay
(A13)
Remove
screw with
cap. washer
(1 place)
Remove
screws
(7 places)