12. Installing the CO2 (A23) Modules (continued)
12.7 Place the OEM bracket into rear case and
secure with the three new mounting screws
(3207337-312), torque to 6.8 inlb.
12.8 If present, connect the NIBP hose from the
module to the inlet fitting. Route excess tubing
as shown. Ensure that the NIBP tubing is
routed through the hole in the bracket cover.
12.9 Connect CO2 module tube to CO2 inlet tube.
NOTE:
The Nafion tube can cause pinching if
pushed back into the CO2 module area. Keep
the tube from the module stretched out while
connecting the inlet tube.
12.10 Connect CO2 inlet cable (W28) to CO2 adapter
cable (W30).
12.11 Connect the exhaust tube to the rear case
outlet connector.
Connect Tubing
NIBP Option
Screw
3207337-312
(3 places)
T
O
R
Q
U
E
T
O
R
Q
U
E
T
O
R
Q
U
E
6.8 In-Lb
CO2 Inlet Cable
Connection
Connect Tubing
CO2 Option
Connect CO2
exhaust tubing