13. Installing the OEM PCB (A06) (continued)
13.8 Connect the power/contact cable (W05) to the
power PCB J12.
13.9 Connect the power/system cable (W01) to the
power PCB J17.
NOTE:
The W01 - power/system PCB cable is
under the OEM shield.
13.10 Install the OEM shield (3208298-000).
13.11 Connect the W02 - power/therapy PCB cable
(3009726) to J8 on the power PCB.
13.12 Route the left side of the OEM shield under the
power/therapy cable J8 connector and over the
top of J11, J13 and J15 connectors on power
PCB.
PWR/Therapy
Cable
3009726-05 (W02)
OEM Shield
3208298
Power/System
Cable 3207692
(W01)
Power/Contact
Cable 3207692
(W05)
Route CO2 Wire
Harness As
Shown