9.
Removing the Parameter Bezel
9.1
Disconnect the power/contact PCB (W05) from
the power board J12.
9.2
Remove the power/system cable (W01) from
the power board J17.
9.3
If present, carefully disconnect the SpO2
connector flex cable (W22) (3206995-004) from
J1 of the SpO2 PCB.
CAUTION
POSSIBLE COMPONENT DAMAGE
The SpO2
connector flex cable can be damaged during
disassembly. If SpO2 connector is damaged, replace
the SpO2 flex cable using the MASIMO Sp02 Panel
Mount Cable Repair Kit (MIN 3305431-026) kit.
9.4
Remove the parameter bezel label from the
bezel.
9.5
Remove without CO2 spacer label (3207407-
000) from bezel.
9.6
Remove the two bezel mounting screws.
Discard the screws.
9.7
Lift out the parameter bezel (3206723-005)
from the rear case.
Remove
Screws
(2 places)
Disconnect J12
PWR/Contact
cable (W05)
Disconnect J17
PWR/System
cable (W01)
Remove
Parameter
Bezel Label
Remove W/O
CO2 Spacer
Label