Title
LIFEPAK 15 Monitor Defibrillator CO2 Upgrade
Page
27 of 33
Doc Type
Service Process Document
Doc ID
3308217
Doc Rev
A
Doc State
Released
State Date
09/08/2011
15. Installing the System (A01)/Therapy (A04) PCB
Assembly (continued)
15.9 Connect the W20 - flex cable from the therapy
PCB J51 to the biphasic PCB at J106.
15.10 Line up the system PCB with the inserts in the
rear case, and secure with seven new screws
(202253-761). Torque 6.8 in-lb.
15.11 Capture the ground strap from the EMI Shield
with one new screw with washer (3207337-
312). Torque 6.8 in-lb.
NOTE:
The screw in the ECG shield can be
mis-positioned. Check to ensure that the
screw is installed and torqued.
Screws
202253-761
(7 places)
T
O
R
Q
U
E
T
O
R
Q
U
E
T
O
R
Q
U
E
6.8 In-Lb
Screw w/
Captive Washer
3207337-312
Connect J106
BTE Flex (W20)
1