Changing the Power Supply/Metal Chassis Assembly
4 Disassembly/Reassembly
35
Step 3: Removing the Receiver Board
1
First remove the MIF board (see Step 1).
2
Remove the four screws securing the receiver board to the chassis using a T-10 Torx driver.
3
Gently lift the front right hand corner of the receiver board, at the same time pulling it forward.
4
Next, lift the rear left hand corner of the board so that the board is clear of the metal screw fixing post.
Screw locations
Screw fixing post