Removing and Replacing the Transducer Battery
CAUTION
NEVER immerse a transducer in liquid if the battery drawer is open or has been removed. Check the
condition of the drawer before refitting, and verify that it is fully closed before use.
Correct use
Recycle according to local laws
Incorrect use
x1
x1
Use only this tool to open drawer
M2720-64001 Battery Replacement Kit Contents:
x1
Central
insertion point
(unmarked)
Old battery
New battery
Old battery
New battery
Old battery
Silicone grease inside
gasket
continued > >