Testing Transducers
7 Troubleshooting
73
3
The ultrasound transducer contains seven piezoelectric crystals. Basic functioning of each can be verified
by holding a flat bottomed pencil or similar above each crystal and moving it up and down as shown.
Figure 4 Testing an Ultrasound Transducer Using a Pencil
A sound should be heard for each crystal tested. The pencil should be held two to three centimeters from
the transducer surface when the test is carried out.
Figure 5 Position of Crystals in an Ultrasound Transducer
4
A sound should also be heard when the transducer is moved back and forth over a solid surface, or the
hand as shown below.
Figure 6 Checking an Ultrasound Transducer
psm18sca.tif
ustrancs.hpg
Crystals
pop21sca.tif