Action Settings
8 Changing Settings
83
Action Settings
The action settings can be changed
only
on the base station. To change the action settings, follow the step-
by-step instructions.
Displaying the Software Revision
Clearing the Factory Information Code (FIC) Log
Step
Action
Base station display
looks like...
1
Press the two arrow keys
simultaneously. The bed label
blinks, the two-digit display goes blank.
2
Press
twice to enter the ‘A’ settings. ‘A’ flashes in the
display.
3
Press
. ‘A1’ appears, and the ‘1’ blinks.
4
Press
once to increment ‘A’ setting to 2.
5
Press
. Two-digit display shows
the software revision, in the format
A.XX.XX, as a sequence. For example,
revision A.01.00 is shown in the
following way. This cycles for 30
seconds before the display reverts
back to normal.
Transducer display (if
docked)...
Base station display
shows sequence
1
2
3
Step
Action
Display looks like...
1
Press the two arrow keys
simultaneously. The bed label
blinks, the two-digit display goes blank.
2
Press
twice to enter the ‘A’ settings. ‘A’ flashes in the
display.