10 Specifications
Cables
90
Cables
ECG Frontend
Type
Two Lead ECG
Input Impedance
> 10M
@ 35 Hz
CMRR
> 110 dB
(with 51.1 kW || 47nF imbalance @
line frequency)
Noise
< 4
Vp @ 25 k
input impedance
Contact Potential
± 500 mV
Inop Amplitude at open LA/ RA contacts
60 to 90 mV
Inop Auxiliary Current
< 100 nA
Input Voltage Range ECG
20
Vpp to 4 mVpp (66dB)
Input DC tolerance
± 400 mV
Dielectric Strength
1500 Vrms
Frequency passband
0.7 to 80 Hz
Defibrillator Protection
None
ESU Protection
None
AUX Frontend
Communication Protocol
Serial, 1 start bit, 1 stop bit, 8 data bits,
no parity
Serial Communication Voltage Levels
Unipolar 3 V
Receive: mark = 0V, space = ~3 V
Transmit: mark = 0 V, space = high
impedance (requires pull up resistor)
Communication Speed
fixed 1200 Baud
Max. Output Current for External Devices
100 mA electronically limited
Output Voltage for External Devices
3 V ± 2%
Cable Type
Option Number
Part Number
Length
Avalon Fetal Monitor Interface Cable
(for monitors with software revision
D.00.00 or greater)
M2720A option K30
M2731-60001
2.5 m approx.
Series 50 Interface Cable
M2720A option K50
M2720-61603
1.6 m approx.
Service Tool Cable
-
M1360-61675
3.0 m
Power Cable
-
Country dependent
2.4 m
Frontends
Summary of Contents for Avalon CTS M2720A
Page 2: ...ii ...
Page 32: ...3 System Interfaces Compatible Fetal Monitors 26 ...
Page 66: ...6 Preventive Maintenance and Safety Battery Exchange 60 ...
Page 92: ...9 Upgrades 86 ...
Page 102: ...10 Specifications Essential Performance 96 ...
Page 104: ...A Removing and Replacing the Transducer Battery 98 Wipe off excess silicone grease ...
Page 108: ...C Avalon CTS Frequency Table 102 ...