Essential Performance
10 Specifications
95
The second table identifies Essential Performance in the presence of “transient” interference:
• Electrostatic Discharge (ESD)
• Electrical Fast Transients/Bursts
• Surges
• Voltage interruptions
• Electro Surgery (ESU)
Parameter
Essential Performance
General
No interruption or cessation of current operating mode (e.g., no reboot, display is functional,
etc.)
No spontaneous operation of controls (e.g., no activation of touch screen controls without
user interaction)
ECG
Measurement of the heart rate within the specified error limits and over the specified
measurement ranges
Alarming on heart rate limit violation within specified delay time.
Toco Extern
Measurement of the external Toco within specified error limits and over the specified
measurement range
Ultrasound
Measurement of the fetal heart rate within the specified error limits and over the specified
measurement ranges
Alarming on heart rate limit violation within specified delay time.
Parameter
Essential Performance
All
After electrostatic discharge, fast transients/bursts, surges and electro surgery interference, the
equipment will return to previous operation mode within 10 seconds without loss of any stored
data.
After voltage interruptions the equipment returns to previous state without operator intervention
and loss of data.
Summary of Contents for Avalon CTS M2720A
Page 2: ...ii ...
Page 32: ...3 System Interfaces Compatible Fetal Monitors 26 ...
Page 66: ...6 Preventive Maintenance and Safety Battery Exchange 60 ...
Page 92: ...9 Upgrades 86 ...
Page 102: ...10 Specifications Essential Performance 96 ...
Page 104: ...A Removing and Replacing the Transducer Battery 98 Wipe off excess silicone grease ...
Page 108: ...C Avalon CTS Frequency Table 102 ...