Function Settings
8 Changing Settings
81
Selecting the Channel Frequency Spacing
See also page 54. Consult Philips Support before changing the channel frequency spacing.
The Service Support Tool is required to set the base frequency. The channel frequency spacing can be set
using the Service Support Tool (see
Service Support Tool Help
for details), or on the base station as follows:
4
Press
once to increment ‘F’ setting to 2.
5
Press
. Two-digit display shows the current setting
(2.0=automatic channel search, 2.1=fixed frequency)
6
Press
or
to change the setting.
7
Press
to accept the new setting and return to normal
operation.
Step
Action
Display looks like...
Step
Action
Display looks like...
1
Press the two arrow keys
simultaneously. The bed label
blinks, the two-digit display goes blank.
2
Press
three times to enter the ‘F’ settings. ‘F’ flashes in the
display.
3
Press
. ‘F1’ appears, and the ‘1’ blinks.
4
Press
twice to increment ‘F’ setting to 3.
5
Press
. Two-digit display shows the current setting. (Channel
spacing can be set between 0 and 8 inclusive.)
3.0 = default spacing
3.1 = 1 x 12.5 kHz = 12.5 kHz
3.2 = 2 x 12.5 kHz = 25 kHz
3.n = n x 12.5 kHz and so forth
Summary of Contents for Avalon CTS M2720A
Page 2: ...ii ...
Page 32: ...3 System Interfaces Compatible Fetal Monitors 26 ...
Page 66: ...6 Preventive Maintenance and Safety Battery Exchange 60 ...
Page 92: ...9 Upgrades 86 ...
Page 102: ...10 Specifications Essential Performance 96 ...
Page 104: ...A Removing and Replacing the Transducer Battery 98 Wipe off excess silicone grease ...
Page 108: ...C Avalon CTS Frequency Table 102 ...