Regular Preventive Maintenance
6 Preventive Maintenance and Safety
55
WARNING
Do not connect any devices that are not supported as part of a system.
System Example
This illustration shows a system where both the medical electrical equipment and the non-medical
electrical equipment is situated at the patient’s bedside.
WARNING
Any non-medical device placed and operated in the patient’s vicinity (within a distance of 1.5m from the
patient’s bed) must be powered via an approved separation device (for example, an isolation transformer).
If the personal computer (or any other non-medical electrical device) is situated outside the medically used
room, you must take measures to reduce leakage currents, such as providing an additional protective earth,
a non-conducting enclosure, or a separation device.
We highly recommend using a separation device whenever you connect non-medical electrical equipment.
Regular Preventive Maintenance
The care and cleaning requirements that apply to the Avalon CTS and the monitoring accessories are
described in the
Instructions for Use
. This section details the periodic maintenance recommended for the
base station, transducers and accessories.
Mechanical Inspection
All rear panel connections must be tight. Check all exterior housings for cracks and damage. Check the
condition of all external cables, especially for splits or cracks and signs of twisting. If serious damage is
evident, the cable should be replaced immediately. Check that all mountings are correctly installed and
secure. Refer to the instructions that accompany the relevant mounting solution. Check that the protective
earth cable is properly and securely connected.
Non-Medical Devices
Medical Devices
Fetal Monitor
Avalon CTS
Personal
Computer
Key:
Power cables:
Data cables:
Separation
Device
To Purchase, Visit
Summary of Contents for Avalon CTS M2720A
Page 2: ...ii ...
Page 32: ...3 System Interfaces Compatible Fetal Monitors 26 ...
Page 66: ...6 Preventive Maintenance and Safety Battery Exchange 60 ...
Page 92: ...9 Upgrades 86 ...
Page 102: ...10 Specifications Essential Performance 96 ...
Page 104: ...A Removing and Replacing the Transducer Battery 98 Wipe off excess silicone grease ...
Page 108: ...C Avalon CTS Frequency Table 102 ...