8 Changing Settings
Function Settings
80
Function Settings
To change the function settings manually, follow the step-by-step instructions.
Bed Label Appearance
The bed label can be set to appear as two-digit numeric characters between 00 and 99 (factory default), or
as a mix of one alphabetic and one numeric character (for example, A1, B5, C9 and so forth). To change
the way the bed label is displayed:
Enabling/Disabling Fixed Frequency
The Service Support Tool is required to set the base frequency. Fixed frequency selection can be enabled
on the base station, as follows.
Step
Action
Display looks like...
1
Press the two arrow keys
simultaneously. The bed label
blinks, the two-digit display goes blank.
2
Press
three times to enter the ‘F’ settings. ‘F’ flashes in the
display.
3
Press
. ‘F1’ appears, and the ‘1’ blinks.
4
Press
again. Two-digit display shows the current setting
(1.0=numeric characters only, 1.1=mixed alphabetic and numeric
characters)
5
Press
or
to change the setting.
6
Press
to accept the new setting and return to normal
operation.
Step
Action
Display looks like...
1
Press the two arrow keys
simultaneously. The bed label
blinks, the two-digit display goes blank.
2
Press
three times to enter the ‘F’ settings. ‘F’ flashes in the
display.
3
Press
. ‘F1’ appears, and the ‘1’ blinks.
Summary of Contents for Avalon CTS M2720A
Page 2: ...ii ...
Page 32: ...3 System Interfaces Compatible Fetal Monitors 26 ...
Page 66: ...6 Preventive Maintenance and Safety Battery Exchange 60 ...
Page 92: ...9 Upgrades 86 ...
Page 102: ...10 Specifications Essential Performance 96 ...
Page 104: ...A Removing and Replacing the Transducer Battery 98 Wipe off excess silicone grease ...
Page 108: ...C Avalon CTS Frequency Table 102 ...