10 Specifications
Transducers
88
Transducers
Antenna
Input Impedance
50
Water Ingress Protection Code
IP X1 (protection only against vertically falling water drops)
Monitor Interface
Toco Output
Accuracy
± 0,5% per 100 mmHg
(not including transmitter)
Offset
± 5 Units
(not including transmitter)
Range
0 to 4V
Voltage Range
US Voltage range
4mVpp to 4Vpp
ECG Voltage range
0.1Vpp to 4Vpp
Transducer Specifications
General
Shock Resistance
Withstands a 1m drop to concrete surface with possible cosmetic damage
Usability Underwater
0.5m/1.64 ft
Water Ingress Protection Code
IP 68 (0.5m/1.64 ft immersion for 5 hours)
Dimensions and Weight
Size (diameter)
< 10 cm/3.94 in
Weight
140 g/4.8 oz.
Battery
Type
Lithium Ion
Capacity
> 16 hours (> 700mAh)
Life
> 500 charge/discharge cycles
(with new battery, at 25°C/77°F)
Transducer Storage Time
1 year at 25°C/77°F
(battery full)
1 month at 25°C/77°F
(battery empty)
Recharging Time
100% charged
3h
66% charged
1h
Forced discharge current
approx. 45mA
Discharge leakage current in
shutdown @ Vbat = 3V, 25°C
20
A (typical), 70
A (max)
Degree of Protection Against
Electrical Shock
Type CF
RF Unit
Nominal RF Output Power
-12 dBm ± 6 dB ERP
Carrier Frequency Range
See Frequency Bands
Minimum Frequency Band Span
Per Option
10 MHz
Channel spacing
25kHz or 12.5kHz
Base Station Specifications
To Purchase, Visit
Summary of Contents for Avalon CTS M2720A
Page 2: ...ii ...
Page 32: ...3 System Interfaces Compatible Fetal Monitors 26 ...
Page 66: ...6 Preventive Maintenance and Safety Battery Exchange 60 ...
Page 92: ...9 Upgrades 86 ...
Page 102: ...10 Specifications Essential Performance 96 ...
Page 104: ...A Removing and Replacing the Transducer Battery 98 Wipe off excess silicone grease ...
Page 108: ...C Avalon CTS Frequency Table 102 ...