ALARM PARAMETERS
Setting door alarm delay
This function delays the activation of the door alarm for the preset amount of time.
Settable range: 1 – 30 minutes
Factory setting: 2 minutes
1. On the home screen, press the menu key (MENU/HOME).
▶
The left side of the display changes to the Menu screen.
2. On the Menu screen, move the cursor to Alarms & Controls [Fig. 1]
using the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The display changes to the Alarms & Controls screen.
3. On the Alarms & Controls screen, move the cursor to Other Alarm Set
[Fig. 2] using the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The display changes to the Other Alarm Set screen.
4. When the cursor in the Other Alarm Set screen is on Door Alarm
Delay [Fig. 3], press the enter key (ENTER).
▶
The right side of the display changes to the Door Alarm Delay screen,
and the current setting value (02) is displayed
[Fig. 4].
5. Move the cursor using the left/right keys (
◁▷
) and select the digit you
want to change, and then use the up/down keys (
△▽
) to change the
number.
6. Press the enter key (ENTER).
▶
The value is confirmed and the display returns to the Other Alarm Set
screen.
*If the entered setting value is out of the settable range, the following
message is displayed on the right side of the display. Press any key on
the control panel to change to the
Door Alarm Delay
screen [Fig. 4].
7. Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90 seconds has passed without any key
operation.
Note:
When the incubator is recovered from the alarm state within the set door alarm delay time, the
buzzer doesn’t sound after the elapse of the set door alarm delay time.
<Left side of the display> [Fig. 3]
Other Alarm Set
Door Alarm Delay
Ring Back Delay
Remote Alarm
Door Alarm Delay
02
min
1min – 30min
<Right side of the display> [Fig. 4]
<Left side of the display> [Fig. 1]
Menu
Set Value
Data Log
Lock
Alarms & Controls
Door Alarm Delay
Invalid Value
Press Any Key
1min – 30min
<Right side of the display>
<Left side of the display> [Fig. 2]
Alarms & Controls
Temp/Gas Alarm Set
Other Alarm Set
UV Setting
Gas Supply Setting
Date & Time
Keypad Click Set
▼