39
BASIC PARAMETERS
Setting high limit temperature alarm
The high limit temperature alarm is different from the temperature alarm (page 61), and it is independent
temperature alarm. When the chamber temperature exceeds the temperature set for the high limit
temperature alarm, this alarm activates.
The high limit temperature for the normal incubator operation can be set as follows. The incubator
automatically starts operation using these settings after power-on.
Settable range: 20.0
o
C – 53.0
o
C
Factory setting: 53.0
o
C
Notes:
ꞏ
Set the temperature of the high limit temperature alarm after the chamber temperature becomes stable at the
set value.
ꞏ
Set the high limit temperature alarm to at least 1
o
C higher than the temperature set for the chamber.
1. On the home screen, press the menu key (MENU/HOME).
▶
The left side of the display changes to the Menu screen.
2. When the cursor on the Menu screen is on Set Value [Fig. 1], press
the enter key (ENTER).
▶
The display changes to the Set Value screen.
3. On the Set Value screen, move the cursor to Overtemp [Fig. 2] using
the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The right side of the screen changes to the Overtemp screen, and the
current setting value (53.0) is displayed [Fig.3].
4. Move the cursor using the left/right keys (
◁▷
) and select the digit you
want to change, and then use the up/down keys (
△▽
) to change the
number.
5. Press the enter key (ENTER).
▶
The value is confirmed and the display returns to the Set Value
screen.
*If the entered setting value is out of the settable range, the following
message is displayed on the right side of the display: Press any key on
the control panel to change to the Overtemp screen [Fig. 3].
6. Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90 seconds has passed without any key
operation.
Overtemp
Invalid Value
Press Any Key
20.0°C – 53.0°C
<Right side of the display>
<Right side of the display> [Fig. 3]
Overtemp
53.0
°C
20.0°C – 53.0°C
<Left side of the display> [Fig. 2]
<Left side of the display> [Fig. 1]
Menu
Set Value
Data Log
Lock
Alarms & Controls
Set Value
Temperature
:
37.0
o
C
CO
2
ON/OFF
: ON
CO
2
Setting
: 5.0%
O
2
ON/OFF
: ON
O
2
Setting
: 5.0%
Overtemp
: 53.0
o
C