SPECIFICATIONS
Product name
O
2
/CO
2
Incubator MCO-50ML
External dimensions
W 480 mm x D 550 mm x H 585 mm (W 18.9 in. x D 21.7 in. x H 23.0 in.)
Internal dimensions
W 370 mm x D 363 mm x H 385 mm (W 14.6 in. x D 14.3 in. x H 15.2 in.)
Interior volume
50 L (1.77 cu.ft.)
Exterior Painted
steel
(Back cover has no paint)
Interior
Stainless steel containing copper
Outer door
Painted steel
Inner door
Tempered glass
Trays
2 trays made of stainless steel containing copper
W 353 mm x D 308 mm x H 12 mm (W 13.9 in. x D 12.1 in. x H 0.47 in.)
Maximum load: 7 kg (15 lbs.)/tray
Access port
Inner diameter: 30 mm (1.18 in.); on the back side
Insulation Styrene
AcryloNitrile
copolymer
Heating system
DHA system (heater air jacket system)
Heater 165
W
Humidifying system
Natural evaporation with humidifying pan
Temperature controller
PID control system
Temperature display
Digital display
CO
2
controller
PID control system
CO
2
density display
Digital display
O
2
controller
PID control system
O
2
density display
Digital display
Air circulation
Fan-assisted
Gas line filter
0.01
m, Efficiency: 99.99 % or higher
Alarms
Temperature alarm, CO
2
density alarm, O
2
density alarm,
High-limit temperature alarm, CO
2
/O
2
/N
2
gas alarm, heater alarm, fan stop alarm
Remote alarm contacts
Allowable contact capacity: DC 30 V, 2 A
*1
CO
2
inlet connection
Soft Polyurethane tube can be connected.
(ID
4 mm, OD 6 mm (ID 0.157 in. ,OD 0.236 in.))
CO
2
inlet pressure
0.03 MPa(G)–0.1 MPa(G) (0.3 kgf/cm
2
(G)–1 kgf/cm
2
(G), 4.4 psi(G)–14.5 psi(G))
O
2
inlet connection
Soft Polyurethane tube can be connected.
(ID
4 mm, OD 6 mm (ID 0.157 in. ,OD 0.236 in.))
O
2
inlet pressure
0.05 MPa(G)–0.1 MPa(G) (0.5 kgf/cm
2
(G)–1 kgf/cm
2
(G), 4.4 psi(G)–14.5 psi(G))
Weight
46 kg (101 lbs.)
Accessories
1 power supply cord cover plate, 2 trays, 3 gas tubes, 1 humidifying pan,
1 Gas injection nozzle, 1 Gas injection nozzle tube, 3 conversion joints,
1 removable power supply cord for UK,
1 removable power supply cord for EU countries