65
ALARM PARAMETERS
Setting O
2
density alarm
When the O
2
density goes out of the preset range (the set O
2
density ± the set O
2
density alarm value), the
alarm activates.
Settable range: ±0.5 % – ±5.0 %
Factory setting: ±1.0 %
1. On the home screen, press the menu key (MENU/HOME).
▶
The left side of the display changes to the Menu screen.
2. On the Menu screen, move the cursor to Alarms & Controls [Fig. 1]
using the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The display changes to the Alarms & Controls screen.
3. When the cursor in the Alarms & Controls screen is on Temp/Gas
Alarm Set [Fig. 2], press the enter key (ENTER).
▶
The display changes to the Temp/Gas Alarm Set screen.
4. On the Temp/Gas Alarm Set screen, move the cursor to O
2
Alarm [Fig.
3] using the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The right side of the display changes to the O
2
Alarm screen, and the
current setting value (±1.0) is displayed [Fig. 4].
5. Move the cursor using the left/right keys (
◁▷
) and select the digit you
want to change, and then use the up/down keys (
△▽
) to change the
number.
6. Press the enter key (ENTER).
▶
The value is confirmed and the display returns to the Temp/Gas Alarm
Set screen.
*If the entered setting value is out of the settable range, the following
message is displayed on the right side of the display: Press any key on
the control panel to change to the O
2
Alarm screen [Fig. 4].
7. Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90 seconds has passed without any key
operation.
<Left side of the display> [Fig. 2]
Alarms & Controls
Temp/Gas Alarm Set
Other Alarm Set
UV Setting
Gas Supply Setting
Date & Time
Keypad Click Set
▼
<Right side of the display> [Fig. 4]
O
2
Alarm
±1.0
%
±0.5% – ±5.0%
<Left side of the display> [Fig. 1]
Menu
Set Value
Data Log
Lock
Alarms & Controls
O
2
Alarm
Invalid Value
Press Any Key
±0.5% – ±5.0%
<Right side of the display>
<Left side of the display> [Fig. 3]
Temp/Gas Alarm Set
Temp Alarm
Temp Alarm Delay
CO
2
Alarm
CO
2
Alarm Delay
O
2
Alarm
O
2
Alarm Delay