OPERATION/ALARM LOG
Displaying operation log
The product's operation history can be graphically displayed on the display.
Furthermore, it is possible to export these records to USB flash drive.
Graph display setting: Internal temperature, CO
2
density, O
2
density,
and the door opening/closing records.
Initial display: Internal temperature display
1. On the home screen, press the menu key (MENU/HOME).
▶
The left side of the display changes to the Menu screen.
2. On the Menu screen, move the cursor to Data Log [Fig. 1] using the
up/down keys (
△▽
), and press the enter key (ENTER).
▶
The display changes to the Data Log screen.
3. On the Data Log screen, move the cursor to Data Log Chart [Fig. 2]
and press the enter key (ENTER).
▶
The right side of the display changes to the graph screen, and the
current setting value (graph showing the temperature in the chamber) is
displayed [Fig. 3].
4. With Temp on the upper left of the screen highlighted, press the enter
key (ENTER).
▶
The display changes to the Data Select screen and the current setting
value (TEMP) is displayed [Fig. 4].
5. Use the up/down keys (
△▽
) to select the data item to be displayed in
a graph.
Each time the up/down key is pressed, the display changes between
Temp, CO
2
, O
2
, and Door screens.
Temp: the record of temperatures in the chamber
CO
2
:
the record CO
2
densities in the chamber
O
2
:
the record of O
2
densities in the chamber
Door: the door opening/closing record
6. Press the enter key (ENTER).
▶
A graph of the data is displayed [Fig. 5].
*Door opening/closing record graph display [Fig. 6]
OP: Door opened
CL: Door closed
<Right side of the display> [Fig. 4]
Data Select
TEMP
Temp / CO
2
/ O
2
/ Door
<Left side of the display> [Fig. 2]
Data Log
Data Log Chart
Data Log Export
Data Log Setting
Alarm
Alarm Export
<Left side of the display> [Fig. 1]
Menu
Set Value
Data Log
Lock
Alarms & Controls
<Right side of the display> [Fig. 3]
-
‐
-
‐
-
Temp
01/Jan/19 EXP.
50°C
40°C
30°C
0
12 24
<Right side of the display> [Fig. 5]
-
‐
-
‐
-
Temp
01/Jan/19 EXP.
40°C
30°C
20°C
0
12 24
<Right side of the display> [Fig. 6]
Door
01/Jan/19 EXP.
OP
―
CL
―
0
12 24