DECONTAMINATION (OPTION)
*If the H
2
O
2
decontamination is
terminated or ends in error, the
error number, error code, and error
message is displayed after
H
2
O
2
decontamination [Fig. 14].
You can see the error information
by pressing the left/right keys.
*If the Cancel key (CANCEL) is
pressed to stop the decontamination
process during Step 1 or Step 2, a
confirmation screen is displayed on
the right side of the display [Fig. 15].
Change the setting to YES using the
up/down keys (
△▽
), and then press
the Enter key (ENTER).
Caution: H
2
O
2
decontamination
cannot be stopped during Step 3
(UV resolve).
If decontamination is canceled
during Step 1, the screen in Fig.
16 is displayed. Open the outer
door and inner door, remove the
power connector of the H
2
O
2
generator from the connector of
the main unit, and then remove
the H
2
O
2
generator to complete
the decontamination work.
If decontamination is canceled
during Step 2, the screen in Fig.
17 is displayed, Step 3: UV
resolve starts, and then the
screen in Fig. 18 is displayed
after the H
2
O
2
vapor has been
broken down. Open the outer
door and inner door, remove the
power connector of the H
2
O
2
generator from the connector of
the main unit, and then remove
the H
2
O
2
generator to complete
the decontamination work.
Step 3 : UV Resolve
90
min
Power Failure
H
2
O
2
Decontamination
Door : Look
Step 1
Step 2
UV : ON
Step 3
◀
05/10
▶
Err35 :
<Left side of the display>
<Right side of the display> [Fig. 14]
<Right side of the display> [Fig. 15]
Decon Aborted.
Please Remove
H
2
O
2
Unit.
H
2
O
2
Decontamination
Temp.
Door : Closed
UV : OFF
37.0
Abort?
NO
YES / NO
Step 3 : UV Resolve
90
min
H
2
O
2
Decontamination
Door : Look
Step 1
Step 2
UV : ON
Step 3
Decon Aborted.
Please Remove
H
2
O
2
Unit.
H
2
O
2
Decontamination
Temp.
Door : Closed
UV : OFF
48.0
<Left side of the display>
<Right side of the display> [Fig. 16]
<Left side of the display>
<Right side of the display> [Fig. 17]
<Left side of the display>
<Right side of the display> [Fig. 18]
<Left side of the display>
H
2
O
2
Decontamination
Door : Look
Step 1
Step 2
UV : OFF
Step 3